background image

P84432 G 

Sheet 3 of 6

 

 

When calculating the total currents:  Use Table 1/1A to determine the highest value of “RMS Current” or “Average Current” for 
an individual horn (across the expected operating voltage range of the horn), then multiply these values by the total number of 
horns; be sure to add the currents for any other appliances, including audible signaling appliances, powered by the same source 
and include any required safety factors. 

 

If the peak current exceeds the power supplies’ peak capacity, the output voltage provided by the power supplies may drop 
below the listed voltage range of the appliances connected to the supply and the voltage may not recover in some types of power 
supplies.  For example, an auxiliary power supply that lacks filtering at its output stage (either via lack of capacitance and/or 
lack of battery backup across the output) may exhibit this characteristic. 

 

WARNING:  THE HS-24 APPLIANCE MUST BE FIELD SET TO THE DESIRED TONE AND dBA SOUND 

OUTPUT LEVEL BEFORE IT IS INSTALLED.  THIS IS DONE BY PROPERLY INSERTING JUMPER PLUGS IN 
ACCORDANCE WITH THESE INSTRUCTIONS.  INCORRECT SETTINGS WILL RESULT IN IMPROPER 
PERFORMANCE, WHICH COULD RESULT IN PROPERTY DAMAGE AND SERIOUS INJURY OR DEATH TO 
YOU AND/OR OTHERS.

 

SOUND OUTPUT (SPL) SETTINGS: 
 
NOTE:

  The Code 3 Horn incorporates the temporal pattern (1/2 second on, 1/2 second off, 1/2 second on, 1/2 second off, 1/2 

second on, 1-1/2 off and repeat) specified by ANSI/NFPA for standard emergency evacuation signaling.  

The Code 3 Horn 

should be used only for fire evacuation signaling and not for any other purpose. 
 

Figure 2:  Showing Location of Jumper Plugs 

 

+   STR   -

J3

+   AUD   -

TB1

3

J2

C

O

E

D

J1

H

L

M

P83511 ( )

AH

C1

R5

TOP

3

J2

C

O

E

D

J1

H

L

M

 

Factory setting is on Medium dB and Code 3. 

 

Figure 3:  Jumper Plug Settings for High, Medium, Low dB, 

Code 3 Horn and Continuous Horn Setting. 

 

J1

HIGH

LOW

H

C
O
D
E
3

3

E

D

O

C

SETTING

HORN

SETTING

HORN

J1

J2

J2

J2

M

L

H

M

L

H

M

L

MEDIUM

SETTING

HORN

CODE 3

SETTING

HORN

CONTINUOUS

SETTING

* HORN

 

(Use needle nose pliers to pull and properly insert the jumper plug.) 

 

No jumper plug is needed for continuous horn setting.  However, it is recommended that the jumper plug be retained in the unit 
for future use (if needed) as shown in Figure 3. 

Отзывы: