Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
6/25
Functions
10
20
30
?
+
-
Intensity
mW
cm²
Test / Alarm
>10
mW
cm²
DFSBNPUIFSN
13
12
4
3
2
1
11
10
7
6
5
1
Button illumination On / Off.
2
Button heating On / Off.
3
Label next maintenance.
4
Alarm reset button:
a.) By pushing the button the audible alarm of the safety function >10 mW/cm² is reset and the radiant
warmer proceeds heating at the selected intensity for further 15 minutes.
b.) By pushing the button for approx. 3 seconds the safety function >10 mW/cm² is switched-off which
is confirmed by two short tones and the yellow warning lamp (10) flashes. The safety function is
reactivated by pushing the button again for approx. 3 seconds which is confirmed by one short
tone.
5
Pilot lamp (green) illumination.
6
Pilot lamp (green) heating power.
7
Warning lamp (red):
a.) Flashing during normal operation: The battery capacity for the power failure alarm is below
10 minutes.
b.) Lighting in connection with a permanent alarm tone: Power failure. The power line of the radiant
warmer is interrupted. The alarm can be reset by switching-off the heating (2). When the power line
has been reinstalled, the heating must be switched on again.
8
Idle
9
Idle
10
Warning lamp (yellow) >10mW/cm²:
a.) Lighting: Intensity selection above 10 mW/cm². Increased attention must be given to the patient.
After 15 minutes the radiation intensity is automatically reduced to 10 mW/cm².
b.) Flashing: The safety function >10mW/cm² has been switched-off and the radiant warmer
permanently generates the selected radiation intensity. The patient temperature must be securely
surveyed.
11
Pilot lamp (green) heating power. Lighting when heating power is available. When the lamp lights, the
radiant warmer will heat-up, when it does not light, the radiant warmer will cool-off.
12
Intensity selection display: LED-spot field.
13
Intensity selection knob with 15 increments.