VSF User’s Guide (VSF) Revision AN 14 Sept. 2011
11
Not enough flex.
Wiper may not be effective.
Proper flex.
Wiper maintains contact with
instrument face and
optical window.
Too much flex.
Wiper may cause too much
friction, using excessive
power.
12.
Use the 3/32-in. hex key to tighten the screw to “finger-tight,” then snug an
additional quarter-turn.
13.
Run the instrument to verify operation. The
bio-wiper
™ must rotate 180 degrees to
clear the optics before sampling, and 180 degrees to cover the optics after sampling.
14.
If the wiper needs adjusting, loosen the screw, make any necessary adjustments, and
repeat steps 9 through 13 to ensure the wiper is performing properly.