Wet Labs ac-s Скачать руководство пользователя страница 33

 

 

ac-s User’s Guide (ac-s)  

Revision F 

9 Feb. 2006 

 

25 

reflected back into the water volume by the reflective tube. The light is then collected 
by a diffused large area detector at the far end of the flow tube.  
 
The flow tube uses the internal reflection principle in reflecting light back into the 
water volume. A clear quartz tube is employed. The outer perimeter of the tube is 
enclosed by a thin annular volume of air. Using the Fresnel Equation, one can see that 
with an index of refraction of 1.33 in water and index of refraction of 1 in air, the 
total internal reflection is achieved to 41.7 degrees with respect to the optical axis. 

6.3  

Electronics 

The main electronic components of the ac-s consist of  a power and i/o (input/output) 
board, a microprocessor board, an A/D board, an amplifier board, a motor controller 
board, a dc motor with an integrated position encoder, two incandescent lamps and four 
photodiode detectors. 

6.4  

Signal  Processing 

The purpose of the ac-s signal processing circuitry is to take a raw optical signal and 
make it into a physically meaningful measurement ready for output. Signals from the 
absorption path and attenuation path detectors go through several levels of analog and 
digital processing before they are registered as output from the unit. To understand the 
exact nature of the ac-s signal processing, it is first necessary to better understand the 
primary data sampling. Figure 9 is a timing representation of the ac-s detector signals 
through a single filter wheel rotation. It shows how each of the two filter sections are 
brought in line with the optical path once per revolution.  
 

0

512

1024

1536

2048

Encoder Step

A Ref

A Sig

C Ref

C Sig

Figure 9. 

Signal timing for one filter wheel revolution 

Each trace represents the optical signal from a detector as the filter wheel spins through 
its cycle in relation to the motor encoder position (1 to 2048). During the filter wheel 

Содержание ac-s

Страница 1: ...ac s User s Guide ac s Revision F 9 Feb 2006 Spectral Absorption and Attenuation Meter ac s User s Guide WET Labs Inc P O Box 518 Philomath OR 97370 541 929 5650 www wetlabs com...

Страница 2: ......

Страница 3: ...cannot accept instruments for servicing or repair that are treated with anti fouling compound s This includes but is not limited to tri butyl tin TBT marine anti fouling paint ablative coatings etc P...

Страница 4: ......

Страница 5: ...s and labor and pay for return via 3rd day air shipping in honoring this warranty Shipping Requirements for Warranty and Out of warranty Instruments 1 Please retain the original shipping material We d...

Страница 6: ......

Страница 7: ...and Data Output 11 3 4 Data Format 11 3 5 Post Processing 11 3 6 Temperature dependent Absorption Correction 13 3 7 Scattering Correction 14 4 Calibration and Characterization 15 4 1 Temperature Corr...

Страница 8: ......

Страница 9: ...is different from the absorption tube Its flow chamber is plastic and the two sleeves on the tube are identical This tube installs on the c side of the instrument the side with the identical looking w...

Страница 10: ...4 Figure 4 You should measure the power supply voltage across these two pins 3 Turn the power supply off 4 Connect the pigtail to the instrument Push the connector straight on to avoid damaging the pi...

Страница 11: ...deploy If the values are outside this range the first step is to re clean the instrument and then reapply the CAL file offsets 1 4 Cleaning 1 Remove flow tubes and all O rings Remove the sleeves from...

Страница 12: ...of data and the values should not vary more than 0 005 m 1 over the 10 minute time period 8 Collect data Record a one to two minute file and save data Repeat steps 4 6 until you can collect three data...

Страница 13: ...ons Data from the ac s is transmitted via both RS 232 and RS 485 depending on which sockets are connected to the instrument s bulkhead connector Regardless of the mode of transmission you use you must...

Страница 14: ...o the ac s Applying a small amount of silicone grease or equivalent to the base of the instrument bulkhead makes the plug insertion easier and provides greater assurance of a good seal Use a connector...

Страница 15: ...lowing basic procedures 1 Do not make direct contact between the ac s pressure housing and a metal frame or hose clamp The ac s is available in one of two pressure housing materials Aluminum is used f...

Страница 16: ...o shock The data will be adversely influenced by bubbles dirt or grease in the flow path or misalignment of the flow tubes Make sure that the flow tubes pump tubing and screens are free of dirt and gr...

Страница 17: ...n deck and lash it down Give the cage and instrument a fresh water wash down after every cast If this is not practical wash the instrument at the end of each data collection day Holding a hose at low...

Страница 18: ...WARNING Do not use acetone on the windows It will damage the window holders 2 Wash the windows The windows should be cleaned with dilute soapy water followed by ethyl alcohol and should receive a fin...

Страница 19: ...3 Digital Processing and Data Output The CPU takes multiple samples of both signal and reference channels accumulates them through the sampling period and then averages the values at the end of the s...

Страница 20: ...bration and Characterization We therefore can solve for c or a by c 1 x ln Csig Cref ln N 4 or c 1 x ln Csig Cref 1 x ln N 5 or c ln N x ln Csig Cref x 6 Since the ac s pathlength is a fixed constant...

Страница 21: ...fect must be considered to correct the absorption and attenuation at wavelengths above 700 nm Experimental results show that aT the water specific absorption correction due to temperature at 715 nm is...

Страница 22: ...ust be corrected to obtain accurate absorption values There are several alternatives in performing the scattered light correction The simplest involves subtracting the temperature corrected a715 measu...

Страница 23: ...alf to one degree Celsius through the operational temperature range of the instrument From these values we generate a table of temperature compensation offsets Tn This table is contained in each instr...

Страница 24: ...g lasers used in lectures work very well for this task If you see more than one or two large scatterers in the beam path your water is unsuitable for a clean water calibration WET Labs and researchers...

Страница 25: ...it Effective air calibration requires 1 The instrument s optical path is clean and dry 2 The instrument s optical path is completely shielded from ambient light To perform air tracking we recommend th...

Страница 26: ...tion of 0 002 m 1 or less above 450 nm If you have a substantial slope to the data over the acquisition period your meter may have some residual moisture in the flow tubes or on the windows If the opt...

Страница 27: ...10 cm available Beam cross section 8 mm dia nominal Linearity 99 R2 Output wavelengths 80 90 Resolution 4 nm Precision 450 750nm 0 001 m 1 typ 0 003 max 4 Hz 0 0005 m 1 typ 0 0015 max 1 Hz Precision...

Страница 28: ...rrangements Sea Test Cable Connector Pin or Socket Function 1 Common 2 RS 232 Receive 3 RS 485 4 Voltage in 10 to 35 VDC 5 RS 232 Send to Host 6 RS 485 Voltage supplied to the instrument is internally...

Страница 29: ...this socket Pin and socket assignments for the pump power connector are shown below Figure 5 Pump Cable Connector Pin and Socket Assignments Pump Power Cable Connector Pin or Socket Function 1 Common...

Страница 30: ...pect both tubes and make sure they are free of stains and dust The reflective flow tube for the absorption measurement operates using the principle of internal reflection To maintain its reflective pr...

Страница 31: ...rom both filter sections Each filter covers approximately a 72 degree section of the beam path across the filter wheel The filter wheel rotates at a tightly controlled 8 0 rps such that the shorter wa...

Страница 32: ...ore subject to both scattering and absorptive losses by the water Once through the water path the light passes through another pressure window and then is re focused through a lens upon a receiver det...

Страница 33: ...ion encoder two incandescent lamps and four photodiode detectors 6 4 Signal Processing The purpose of the ac s signal processing circuitry is to take a raw optical signal and make it into a physically...

Страница 34: ...rsects one filter section for about 1 8 revolution before the a beam intersects the other filter This staggering of the filters relative to the beams is the reason for the phase differences between th...

Страница 35: ...required to send the data and requirements on the CPU during sampling the next data sample is not captured until a full even number of revolutions have passed following completion of the data averagin...

Страница 36: ...17 c9 12 d8 14 0a0 32 10 ce 1a 25 14 d2 16 17 12 61 1c b5 16 e8 18 0b0 20 14 08 1f 85 19 2b 1a 5b 15 e7 22 8e 1b ad 1c 0c0 be 17 d8 25 c7 1e 5d 1f 3d 19 ef 29 33 21 3b 21 0d0 e4 1c 17 2c d5 24 48 24...

Страница 37: ...4 ff00ff00 Packet registration 013 2 02D0 Record length of full packet not including chksum 720 015 1 05 Packet Type 03 and above designates an ac s meter 016 1 01 reserved 017 1 53 meter type 53 ind...

Страница 38: ...t two bytes are the A D counts from the pressure sensor circuitry If no pressure sensor is installed these two bytes may be ignored When a pressure sensor is installed the depth in meters can be calcu...

Страница 39: ...senting the number of output wavelengths Next is the data for the scan The values are the averaged signal counts dark count subtracted for each detector at each output bin The counts are 16 bit unsign...

Страница 40: ...32 ac s User s Guide ac s Revision F 9 Feb 2006 Figure 10 ac s system diagram...

Страница 41: ...e oil from escaping from the reservoir into the water However you may occasionally wish to ensure that oil remains in the reservoir on top of the transducer WARNING Never touch or push on the transduc...

Страница 42: ...er this common mode noise will be most apparent To eliminate this noise issue either the laptop the power source or the meter itself needs to be connected to an earth ground i e the ground receptacle...

Страница 43: ...Rhoades B 6 29 04 Update specifications DCR 401 D Stahlke I Walsh C 12 2 04 Update Appendices to reflect optional pressure sensor and common mode noise DCR 439 B Rhoades D 4 11 05 Add ac 9 Protocol D...

Отзывы: