11. TANK HEATING

FXP12M heated ultrasonic cleaner models use silicon rubber 

contact heaters on the outside face of the tank and are 

thermostatically controlled. The liquid level in all cleaners 

should not drop below 80mm from the tank bottom and the 

thermostat setting should not exceed 60°C. Hot water should 

never be poured directly into tank bottom but instead 

be brought to temperature gradually.
Once the cleaner has been turned on according to the 

Operations above, the thermostat can be set to your desired 

temperature. (Note: Heaters come standard wired through 

the timer control to ensure the heater is not accidently left on 

for extended periods of time.)

The aluminium foil test is an economical and consistent 

test for units under 60 Litres to gauge performance. 

FXP8 cleaners over 60 Litres should use a cavitation 

meter for more accurate testing.

 

 

 

 

 

 

 

4

 

12. SERVICING

Unauthorized repairs will void warranty.

Only Unisonics Australia employees should service 

this ultrasonic cleaner.

The circuit board presents high-level voltage. 

Improper testing or repair procedures could result in injury 

to the service person and damage to the product.

All repair requests are supplied as a no obligation free quote.

13. PERFORMANCE TEST:

Please find as follows the testing process for efficiency 

levels in the ultrasonic cleaner.

TESTING PROCEDURE:

Always degas the solution prior to the performance test to allow for the elimination of 

air found in fresh solution. This is achieved by turning the ultrasonics on for 5 minutes.

Testing should always be carried out in a solution only chamber 

(basket and cleaning items removed).

The foil must be suspended in the solution for a duration of 30 seconds.

A consistent pattern of perforation should result with hole sizes varying from one 

to four mm in diameter.

Tested fluid should be disposed of, as aluminium residue will remain and deposit 

as a contaminant.

PLEASE NOTE: 

Inconsistency in foil thickness and manual holding can create 

a degree of error that will vary the results of the foil test.

1.

2.

3.

4.

5

.

WESTLAB
FXP12M
OPERATING

MANUAL

Содержание FXP12M

Страница 1: ...E OF CONTENT 1 Applications 1 2 Specifications 1 3 Construction 1 4 Approvals 1 5 Guidelines 2 6 Do Don t 2 8 General Cleaning 3 9 Installation 3 10 Operation 3 12 Servicing 4 11 Tank Heating 4 13 Performance Test 4 14 Warranty Statement 5 15 Warning 5 Please read this instructions carefully before use ...

Страница 2: ...to avoid excessive spillage of solution when draining fluid Resin coated transducers and circuit board for moisture protection 4 APPROVALS AS NZS 3760 2010 AS NZS 3100 2017 Certificate of Conformity No E990002 C Tick Certificate for inclusion of medical device Class 1 T G A 295 x 150 x 150mm FXP12M Model Timer Mechanical 30min Heating Power Nil Tank Material Pressed 304 Stainless Steel Tank Capaci...

Страница 3: ...ioning covers or auxiliary pans for additional solutions Rinse thoroughly after each cleaning application Keep tank bottom free of silt and dirt accumulation Always drain the basket fully of fluid before leaving the cleaning chamber 7 DON T Run machine without solution in the tank Do not allow liquid level to drop below 80mm from tank bottom Use heaters without a thermostat control Setting for the...

Страница 4: ...r free of dirt contamination and high temperatures Away from possible liquid splatter Easy access for manual adjustment 10 OPERATION Insert the single phase 10A plug into a 240 volt AC outlet Pour cleaning fluid of your choice into the tank Switch the supply mains ON The unit is now ready to operate WESTLAB FXP12M OPERATING MANUAL FOR M MH Turn the Mechanical timer switch to the desired time setti...

Страница 5: ...lia employees should service this ultrasonic cleaner The circuit board presents high level voltage Improper testing or repair procedures could result in injury to the service person and damage to the product All repair requests are supplied as a no obligation free quote 13 PERFORMANCE TEST Please find as follows the testing process for efficiency levels in the ultrasonic cleaner TESTING PROCEDURE ...

Страница 6: ...y fault developing in this machine or its accessories This WARRANTY applies only to installations within the Commonwealth of Australia and its Territories Removal of the original serial numbers voids this WARRANTY in its entirety The MANUFACTURES decision shall be final and binding in respect to any claim or dispute arising out of a defect in any machine parts The MANUFACTURER recognises no other ...

Отзывы: