background image

W-FI Guide 

2

approximately 18 to 24 Gauss, thus an inspector uses the W-FI-20 to ensure his residual 
field is strong enough prior to applying his inspection media (Wet or Dry).  A workpiece 
can be magnetically saturated prior to (or after) inspection, and the inspector must 
measure the field strength and polarity prior employing his demagnetization procedure.  
Therefore, the W-FI-50 is used to measure fields that are too strong for Magnetic Particle 
Inspection. 

 

The W-FI is intended to be disposable, so when the user's 'Quality Control Procedure' 
requires the unit to be recalibrated, it can be simply replaced.  Western does offer 3 
"Verification Fixtures" to measure W-FI accuracy however, as W-FI's can only be 
verified, the user has two choices, either verify it or replace it. This may seem wasteful, 
but W-FI's sell for less than the cost of competitive 'Calibrateable Field Indicators' and 
are priced competitively with Calibration Charges for these Calibrateable units. 

 

Instructions 

 

  Hold the W-FI so the Arrow points toward the surface of the Work Piece.  Laying the 

unit flat on its back is not correct. 

  Do not expose W-FI’s to AC Magnetic Fields. 

  W-FI’s are delicate Instruments, not Hockey Pucks. 

  When performing Demag with a Reversing & Decaying DC Field, pull the W-FI as far 

away as possible.  A typical arm length is satisfactory. 

  “Pinning” the pointer in a strong field will affect the accuracy of a W-FI. 

  Field Orientation can be measured by placing a W-FI flat on it’s back, and rotating it 

until the maximum field is measured.  At the maximum reading, the direction of the 
field is parallel with the instrument pointer (not the arrow). 

 

Verification  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
    
 
 
 

 
 
 
 
 
 

 Verification Fixtures are available for each of the 3 

   models of W-FI’s. 

 To verify the accuracy of a corresponding W-FI, simply 

   place it in the fixture as illustrated, with the arrow lined up
   to the indicator line on the fixture. 

  

Each Verification Fixture has a Certification Decal affixed

   to the front on the Case, Indicating the date of Field 
   Confirmation. 

  

It takes just seconds to verify the accuracy of a W-FI.

 

 

Отзывы: