Step 5: Hold the meter and touch the control solution to
the edge of the sampling end of the test strip. The control
solution will be automatically drawn into the reaction area of
the test strip.
Step 6: The screen will begin to count down. After
8 seconds, the control solution testing result will appear on
UIFNFUFSEJTQMBZTDSFFO"i$UMwøBHXJMMBMTPCFEJTQMBZFE
when results are shown on the screen.
Step 7: Compare the reading on the screen to the control
range printed on the test strip package. If the reading does not fall within the control
range printed on the test strip package, see the chapter “Control Solution Trouble
Shooting“.
Step 8: Remove the used test strip either by pushing the ejection button (if available)
or by pulling it out with your hand. The meter will turn off automatically. The meter will
also time-out after 2 minutes of inactivity.