31
C
ONTROL
SOL
UTION
TESTING
Step 2:
Wait until the flashing blood drop and arrow icons appear on the
meter display screen. The meter will also announce, “Blood glucose
test.Please apply blood onto the test strip” if you have the voice
feature turned on. Use the scroll wheel to enter L1 control solution
testing.
Step 3:
ctl
icon will appear next to the test strip icon and
L1
will appear on the meter dis-
play screen. The meter will also announce “L1 control solution test, please apply
control solution onto the test strip.” If you decide not to do a control solution test,
scroll up or down until
ctl
disappeared from the meter display.
Step 4:
Squeeze a drop of control solution onto a clean, dry, nonabsorbent surface. Do not
apply control solution to the test strip directly from the bottle. Replace the bottle
cap on the control solution bottle immediately after use.
Step 5:
Gently touch the drop of the control solution to the tip of the test strip. The meter
will beep indicating it has received the control solution.