background image

 

2005 Cooper Industries, Inc. 

 

 

6

 

 

 

 

 

S805 

 

 

 

 

 

 

 

 

 

 

 

 

   3/05 

 

 

 

 

 

 

 

 

 

 

CONTACT PYROMETER MAINTENANCE

 

 
    The contact pyrometers have been designed for accuracy, ease 
of use and low cost.  As with any pyrometer, certain guidelines 
should be followed in order to obtain maximum precision of 
measurement. 
 
    Prior to use, it is important that the wire ends (A) be inspected 
and cleaned if necessary. Use a solvent such as alcohol and a stiff 
brush to clean the wires. It is sometimes necessary to use a sharp 
instrument to shave the burned flux off. After about eight hours of 
actual use it will be necessary to either file or clip off the wire ends. 
The wire should then be bent to maintain a gap (B) of .06". This will 
restore the pyrometer to like new condition. This process may be 
repeated as many times as the wire length will allow. 
 

   DISPLAY 

CALIBRATION

 

 
1. 

Set the analyzer rotary selector switch to 

OHMS

2. 

Connect a 100 ohm, 

±

0.1%, resistor from the 

TEST

 receptacle to the 

GND

 

receptacle. 

3. 

Measure voltage from pin 7 (+) to pin 8 (ground) of the ribbon cable (pin 1 of the 
cable is nearest to the battery leads, see figure on page 7) and adjust R3 on the 
display board until display and volt meter readings match, i.e., 1.000V = 100.0 
ohms. 

 

TEMPERATURE CALIBRATION

 

 
1. 

Set the analyzer rotary selector switch to 

TEMP

 and the slide switch to 

°

F

2. 

Connect K-111 or K-181 thermocouple assembly to the 

TEST

 receptacle. 

3. 

Measure room temperature with a calibrated thermometer and record 
temperature T

A

4. 

Adjust R24 until display matches thermometer reading. Set the slide switch to 

°

C

 

and adjust R85 until display reads the 

°

C equivalent of room temperature, i.e., T

A

 

= 72

°

F = 22

°

C. 

5. 

Disconnect thermocouple assembly from the analyzer and set the slide switch to 

°

F

6. 

Using a precision millivolt power supply, input 22.251 mVdc (type K equivalent 
voltage of 1000

°

F) to the 

TEST

 socket. 

7. 

Adjust R84 until display reads 1,000

°

F + (T

A

-32

°

F). For example, if room 

temperature T

A

 = 72

°

F then adjust display to read 1,000

°

F + (72

°

F - 32

°

F) = 

1,040

°

F. Set slide switch to 

°

C

 and adjust R81 until display reads 

°

C equivalent 

of 

°

F input, i.e., 1,040

°

F = 560

°

C. 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Содержание WA2000

Страница 1: ...WA2000 SOLDERING IRON ANAL YZER INSTRUCTION MANUAL ...

Страница 2: ... 0 560 C Accuracy Contact Pyrometer 1 over a ten degree span stability Auxiliary T C 5 of reading Absolute 1 digit Resolution 1 F or 1 C user selectable Sensor CP 01 or CP 02 Contact Pyrometer K 111 K 121 K 181 Type K Thermocouple assembly NOTE Analyzer accuracy only Tolerance of thermocouple and external fixturing must be added Patent pending TIP OHMS TO GROUND Range 0 to 100 Ohms Accuracy 1 of r...

Страница 3: ...rature measurements These assemblies can also be used to monitor active soldering and can measure temperature loading excursions and heater response Although not a requirement for compliance to the MIL STD 2000 specification it does provide a good tool for process control and improvement MEASUREMENT PROCEDURE 1 Set the rotary selector switch to the TEMP position and the slide switch to C or F to s...

Страница 4: ...contact the pyrometer with the iron s tip The iron holder provided with the WA2000 may be used to hold the iron steady Record the analyzer reading V1 4 Turn iron under test on and allow to stabilize With the iron s tip making contact to the pyrometer add additional solder to the tip to increase contact area Allow to stabilize for thirty seconds Record the analyzer reading V2 5 Calculate tip noise ...

Страница 5: ...e the iron s tip to ground resistance IMPORTANT When making measurements always use a clean tinned tip A dirty oxidized tip can cause erroneous measurements Always add additional solder to the tip and pyrometer during measurement to ensure sufficient electrical contact Do not add solder to the thermocouple assemblies RECORDER OUTPUT A buffered output and recorder plug assembly has been provided fo...

Страница 6: ...10 F weekly Visual for damage oxidation and cleanliness daily daily daily Proper operation and performance undefined daily see note 1 daily see note 2 daily ESD per DOD STD 1686 N R N R N R Note 1 Complete soldering within 2 5 seconds Note 2 Complete soldering within 5 seconds Cleanliness only required N R Periodic test not required For confirmation of specifications write to WD6536E 2 CODE 3681 S...

Страница 7: ...GND receptacle 3 Measure voltage from pin 7 to pin 8 ground of the ribbon cable pin 1 of the cable is nearest to the battery leads see figure on page 7 and adjust R3 on the display board until display and volt meter readings match i e 1 000V 100 0 ohms TEMPERATURE CALIBRATION 1 Set the analyzer rotary selector switch to TEMP and the slide switch to F 2 Connect K 111 or K 181 thermocouple assembly ...

Страница 8: ...g of 00 0 on display Allow time for the reading to settle it will take several seconds 4 Connect a precision digital voltmeter to monitor input and adjust signal generator output for a 100 mVAC RMS sinusoidal 60 Hz signal 5 Adjust R77 until the analyzer display matches the voltmeter reading OHMS CALIBRATION 1 Set the analyzer rotary selector switch to OHMS 2 Connect a 100 ohm 0 1 resistor to the a...

Страница 9: ... THE WA2000 Please include model number when ordering these parts from your original supplier Please allow six weeks for delivery of items marked with ITEM DESCRIPTION P N 1 Vinyl Accessory Case C724 2 Ground Cord Adapter 7 3400 3 Contact Pyrometer Large Gauge CP 01 4 Contact Pyrometer Small Gauge CP 02 5 Type K Thermocouple Assembly ETA tip K 111 6 Type K Thermocouple Assembly EPH107 Tip K 121 7 ...

Отзывы: