Directions for Use
Routine Maintenance
45
To enter altitude calibration, press
,
and
simultaneously. The LCD will now
display the Settings Main Menu. When entering the Settings Main Menu, the display will
read as follows:
The cursor will be next to
Altitude - user
check. Press
to enter
Altitude - user
check.
1.
When entering the altitude mode, the display will read as follows:
2.
Select
226 Hz
probe tone.
3.
Place the probe into the 2.0 cm
3
cavity provided with the instrument and check the
cm
3
value against the altitude correction table for accuracy.
4.
If the measured volume is not within the published table value ±/.1 cm
3
, then exit the
altitude mode by pressing the
PROGRAM MODE
button and contact field service.
Providing the measured volume agrees with the published table ±/.1 cm
3
, proceed
with the altitude adjustment.
5.
With the probe still in the 2.0 cm
3
cavity, press the
PROG
button to enter the custom
calibration mode.
CUSTOM
will appear on the fourth line of the display.
6.
The value now displayed in the cm
3
display area is the volume measured and adjusted
to the current altitude. If the value displayed is 2.0cm
3
, the volume is adjusted to the
current site.
7.
If the value is not 2.0 cm
3
, press the
SAVE
button to adjust the volume
measurement to the current altitude.
The measured volume should now read 2.0 cm
3
.
8.
To exit the altitude mode, press the
PAGE
button to return to the Settings main
menu.
9.
Move the cursor using
and
to
Back to Normal
and press the
PAGE
button to return to the Normal mode.
Altitude - user check
Cal modes
Page modes
- Back to normal -
Altitude Mode
ECV 2.0
cm
3
9.99
Standard
Note
CAL MODES and DIAG MODES can only be accessed by Welch Allyn trained
individuals. If these items are selected, "Invalid Selection" will flash on the bottom
right to indicate they are currently disabled.
WARNING
Welch Allyn recommends only trained personnel enter the
Calibration and Diagnostic submenus listed below the Altitude Adjustment in
Calibration mode.
M +
Содержание TM286
Страница 18: ...14 Installation Welch Allyn TM286 Auto Tymp...
Страница 36: ...32 Operation Welch Allyn TM286 Auto Tymp...
Страница 60: ...56 Test Results Welch Allyn TM286 Auto Tymp...
Страница 64: ...60 Computer Interface Welch Allyn TM286 Auto Tymp...
Страница 72: ...68 Standards Specifications and Accessories Welch Allyn TM286 Auto Tymp...