Service Manual
Repair Procedures
67
4.
Remove the Digital board (
5.
Grasp the Silastic
™
gray dome and pull the dome off the button post.
6.
Squeeze the plastic button latches together and then push the button out the front of
the panel.
Replacing the Recharger Board Fuse (F2)
The Recharger board fuse (F2) protects the monitor power supply input at the battery line
to the monitor. If the monitor does not turn on but the BATTERY CHARGING indicator
lights when the ac power adapter is connected, check fuse F2 (3 Amp).
1.
Remove the battery pack (
2.
Open the monitor (
3.
Remove the Recharger board (
“Replacing the Recharger Board”
4.
Replace the fuse (F2 below).
Note
The button dome can be replaced on the post without regard to the orientation of
the notch engraved in the dome.
0696-34
Caution
The fuse is heat sensitive and must be soldered with heat sinks in
place between the fuse body and the solder joint.
Содержание Propaq Encore
Страница 1: ...Propaq Encore Vital Signs Monitor Service Manual...
Страница 14: ...10 Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 50: ...46 Troubleshooting Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 96: ...92 Repair Procedures Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 111: ...Service Manual Technical Overview 107 Cabling Diagrams cabling diag...
Страница 120: ...116 Technical Overview Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 126: ...122 Field Replaceable Units FRUs Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 128: ...124 Manufacturable Test Equipment Welch Allyn Propaq Encore Vital Signs Monitor...
Страница 130: ...126 Dynatech Nevada Patient Simulator Modification Welch Allyn Propaq Encore Vital Signs Monitor...