mVSM Troubleshooting Guide 

 

PERIPHERAL TROUBLESHOOTING GUIDE 

OVERVIEW 

This guide is intended to give an overview of the various problems that can arise when using the 
Bluetooth peripherals associated with the Welch Allyn mVSM platform.  Please also refer to the 
individual peripheral or tablet user manuals. 

COMMON ACTIONS 

Certain actions are common to all peripherals for trouble shooting.  The steps for these actions are 
listed below. 

DISABLE LOCKDOWN ON TABLET 

In many cases a device will be locked down with special software to prevent a patient from tampering 
with the features of the tablet other the Welch Allyn mVSM.  In order to provide resolutions to many 
problems this software will have to be temporarily disabled.   

Tablet with Surelock 

1.

 

Restart the tablet by pressing and holding the power button until the “Device options” menu 
is displayed and then choose “Restart”. 

2.

 

After the device restarts a screen will be displayed say “Please wait while we are loading.”  
Quickly tab on the screen 5 times while this is displayed and a password box should be 
displayed. 

3.

 

Enter the pin “5555” into the password box and choose “OK”. 

4.

 

Choose “Exit SureLock”, when it asks “Are you sure you want to exit SureLock?” choose “Exit”. 
 

RE-ENABLE LOCKDOWN ON TABLET 

After having provided a resolution to a problem the device will need to have the lockdown mode re-
enabled. 

Tablet with Surelock 

1.

 

Open the Applications menu by click on the Apps button on the home screen tablet. 

2.

 

Locate the SureLock icon, this may require swiping right until the icon is displayed. 

3.

 

Press SureLock icon and follow the onscreen instruction for enabling it. 

FIND THE BLUETOOTH MENU ON TABLET 

1.

 

Swipe down from the top of the screen to expose the notification screen.

 

 

2.

 

Touch the gear in the upper right to go to the settings screen 

3.

 

If the tablet has a Connections tab at the top select it. 

4.

 

Find and touch Bluetooth on the left hand side. 

Содержание mVSM

Страница 1: ...Welch Allyn Mobile Vital Signs Monitoring mVSM Troubleshooting Guide For use internally mVSM Troubleshooting Guide 4 7 2015...

Страница 2: ...let 2 Find the Bluetooth Menu On Tablet 2 Peripherals 3 Nonin 9560 Pulse ox 3 Fora IR20b THERMOMETER 5 AND UA 767PBT ci blood pressure monitor 6 AND UC 351PBT ci Scale 8 Clinical Portal Troubleshootin...

Страница 3: ...ayed and then choose Restart 2 After the device restarts a screen will be displayed say Please wait while we are loading Quickly tab on the screen 5 times while this is displayed and a password box sh...

Страница 4: ...consistent reading Welch Allyn mVSM waits until it receives a consistent reading before accepting the data For some patients this can take more time The incorrect peripheral has been associated with t...

Страница 5: ...VSM Troubleshooting Guide 4 Poor patient perfusion Instruct patient to move hands prior to taking measurement Apply a warm washcloth Place extremity on a pillow for support Ambient light Shield sensor...

Страница 6: ...Wait for TiaDoc TM to show in Available Devices 8 Touch TiaDoc TM and when prompted enter the pairing pin 111111 Situation Possible Cause Resolution No data transferred from peripheral device to tabl...

Страница 7: ...ice needs new batteries Replace the batteries in device The incorrect peripheral has been associated with the tablet Log in to the Clinical Portal associated with the tablet and verify that the correc...

Страница 8: ...e blood pressure monitor user manual Observe patient technique and retrain as necessary Correct cuff size Correct placement Correct posture Not talking or moving Arm is at heart level Verify blood pre...

Страница 9: ...tablet Device has become unpaired from tablet Follow pairing instructions to repair the device Device needs new batteries Replace the batteries in device The incorrect peripheral has been associated...

Страница 10: ...and holding on to wall walker chair for support Scale does not clear the carpet Read the scale user manual Retrain the patient on how to take a proper weight Have patient stand in the middle of the sc...

Страница 11: ...tions to the email address associated with the user account If the users email is incorrect in the account the email must be updated by a separate user account that can login Patient Reading Report is...

Страница 12: ...the 5th Step of the patient edit wizard Patient is assigned to a tablet but does not display on the device Patient is set to inactive Check the latest status of the patient in the Status History tab...

Страница 13: ...xpected on Tablets page The tablet is out of cellular range or failed transmission Verify internet connection If device is not connected to the server relocate tablet to be within cellular range If th...

Страница 14: ...the client and training on how to change the batteries is recommended Can I copy paste into the Notes section Yes There is no issue with this though formatting will be lost Can I set up standardized c...

Страница 15: ...ent does not have a scheduled session pending they will be automatically sent to the Home screen Does anybody reimburse for this Yes some state Medicaid programs and private insurances provide reimbur...

Отзывы: