Button Functions

Setup

Assembling the Unit

Your SureTemp Plus thermometer is shipped to 
you with three 1.5V AA batteries installed.

1. Install the oral/axillary probe (blue ejection 

button) or optional rectal probe (red Ejection 
button) by inserting the latching probe 
connector on the end of the coiled cord into the 
unit’s plug receptacle. Ensure that the 
connector snaps into place. The LCD will 
illuminate for approximately 7 seconds and will 
then turn off. 

2. Connect the appropriate removable probe well 

to the thermometer. Note that the cut-out in the 
probe well fits over the probe cord strain relief. 
Ensure that the removable probe well “snaps” 
into position and is completely seated to the 
thermometer. Place the probe into the well.

3. Insert the box of Welch Allyn disposable probe 

covers into the probe cover storage holder. To 
open the box of probe covers, remove and 
discard the tear-away corner. You will see the 
ends of the probe covers in the box. 

Mounting the Unit

Mount the wall-mount bracket in a location such as 
an interior wall or wall divider so that the 
thermometer remains at ambient room 
temperature. Mounting the unit on an uninsulated 
exterior wall, in an area directly hit by sunlight, 
above heat radiators or powered equipment 
(including computers and lights), or below air 
conditioning outlets is not recommended.

Proper mounting of the unit will prevent tangling of 
the probe cord with other equipment at the 
patient’s bedside.

1. Mount the instrument so that all other objects 

are no closer than 4 inches (10.25 cm) from 
the sides and bottom of the unit. 

2. Mount the wall mount a maximum of 48 inches 

(122 cm) from the floor. 

3. Install the wall mount to allow at least 12 

inches 30.5 cm) clearance on top of the 
instrument. This provides ease of removing 
and installing the probe and probe well or 
installing probe covers onto the probe.

Buttons

Procedure

C/F

Press and release the 

C/F 

(Celsius/Fahrenheit) 

button to 

toggle between temperature scales 

any time a temperature is displayed 

on the LCD.

Pulse Timer
(692 Only)

Press and release the 

Pulse Timer

 

button to activate the timer. Audible 

beeps sound at 0, 15, 30, 45, and 

60 seconds. The timer turns off 

automatically after 60 seconds. You 

can stop the timer at any time by 

pressing the 

Pulse Timer

 button or 

by removing the probe from the 

probe well.

Mode 
Selection

Press and release the 

Mode 

Selection

 button after you have 

withdrawn the blue probe from the 

probe well to change the current 

measurement site to Oral, Adult 

Axillary, or Pediatric Axillary. The 

selected measurement site is 

displayed by a flashing body site 

icon. Note that no other 

measurement sites are available 

when using the red probe in the 

Rectal Mode.

Recall

Press and release the 

Recall

 

button when the display is blank to 

recall the last completed predicted 

temperature. The LCD will display 

the recalled temperature for five 

seconds along with the battery 

icon, temperature scale, mode 

selection, and probe-position icon 

(if applicable). 

ৎ㊩

ৎ㊩

ৎ㊩

ৎ㊩

Содержание GS 777

Страница 1: ...GS 777 GS 777 Integrated Wall System Directions for Use Material No 718848...

Страница 2: ......

Страница 3: ...Welch Allyn iii Contents GS 777 1 1 1 1 1 2 2 3 3 3 3 3 SureTemp Plus 5 5 5 6 SureTemp Plus 7 7 7 8 8 8 8 8 9 9 10 10 10 10 11 11 11 11 12 12 12 12 12 13 13 13 13 13 13 13...

Страница 4: ...Cleaning 25 Specifications 26 Environment 26 Additional Technical Information 26 Approvals 26 Warranty 26 SureTemp Plus 27 Symbols and Descriptions 27 Safety and Warnings 28 Introduction 29 SureTemp...

Страница 5: ...mometer and Probe 37 Cleaning the Removable Probe Well 37 Disposal 38 Disposing of Equipment 38 Disposing of Batteries 38 Performance and Technical Specifications 38 Replacement Parts and Accessories...

Страница 6: ...vi Directions for Use Welch Allyn...

Страница 7: ...Welch Allyn 1 GS 777 Welch Allyn 3 5V Welch Allyn Rx only Welch Allyn Welch Allyn WEEE WEEE 2002 96 EC SJ T 11363 2006 BF 1 USB Welch Allyn 2 3 1 3 8 35 5 127...

Страница 8: ...2 Directions for Use Welch Allyn Welch Allyn 3 5V 1 2 3 2 10 4 3 5V 1 2 3 4 1 8 3 5 6 6...

Страница 9: ...llTransformer IEC 60601 1 UL 60601 1 IEC 60601 1 2 CAN CSA C22 2 No 601 1 M90 AS NZS3200 www welchallyn com GreenSeries777WallTransformer II 100 240V 50 60 Hz 0 18A 5V 1 A 5V 1 A 5V 0 5 A www welchall...

Страница 10: ...4 Directions for Use Welch Allyn...

Страница 11: ...Welch Allyn 5 SureTemp Plus REF 690 692 6 000 846 6 036 361 6 971 790 BF AA 1 5 MAX IPX Welch Allyn Welch Allyn Welch Allyn 3 5...

Страница 12: ...6 Directions for Use Welch Allyn Welch Allyn SureTemp Plus Welch Allyn SureTemp Plus Welch Allyn Welch Allyn SureTemp Plus SureTemp Plus Welch Allyn Welch Allyn SureTemp Plus 690 692 SureTemp Plus...

Страница 13: ...Plus SureTemp Plus 690 692 690 Welch Allyn SureTemp Plus Welch Allyn SureTemp Plus SureTemp Plus 690 SureTemp Plus 692 X X X X X X X X X X X X X X X X X X X X X X X 2 3 90 90 1 1 SureTemp Plus 690 Sur...

Страница 14: ...8 Directions for Use Welch Allyn SureTemp Plus 1 5 AA 1 7 2 3 Welch Allyn 1 4 10 25 2 48 122 3 12 30 5 C F C F 692 0 15 30 45 60 60 5 Welch Allyn...

Страница 15: ...Welch Allyn 9 SureTemp Plus 4 6 SureTemp Plus Welch Allyn 7 690 692 3 5 110 0 F 43 3 C 80 0 F 26 7 C 104 0 F 40 0 C 50 0 F 10 0 C...

Страница 16: ...10 Directions for Use Welch Allyn 1 2 3 4 5 6 7 30 8 3 9 10 SureTemp Plus 10 13 17 12 15 18 1 2 3 Welch Allyn Welch Allyn 3 20 Welch Allyn...

Страница 17: ...Welch Allyn 11 1 2 3 4 30 5 5 6 7 SureTemp Plus 10 13 1 2 3 4 1 5 5 8 5 6 30 Welch Allyn Welch Allyn 5 Welch Allyn Welch Allyn...

Страница 18: ...12 Directions for Use Welch Allyn 7 3 8 9 10 3 5 1 2 3 4 5 6 7 8 9 10 SureTemp 692 25 50 100 200 SECURITY COUNT 5 4 3 2 1 10 SEC SECURITY COUNT OFF 3 Welch Allyn Welch Allyn 3 5 3 3 5 60...

Страница 19: ...Welch Allyn 13 5 SEC 30 ON OFF OFF 1 2 3 4 3 AA 1 5 3 AA 1 5 5 6 70 10 CaviCide 70 10 CaviCide AA 1 5...

Страница 20: ...3 690 2 Welch Allyn Inc Welch Allyn Welch Allyn 692 3 690 2 Welch Allyn Welch Allyn Welch Allyn Welch Allyn RMA Welch Allyn Welch Allyn Welch Allyn 1 2 Welch Allyn 3 4 5 50 0 F 104 0 F 10 0 C 40 0 C 5...

Страница 21: ...Welch Allyn 15 1 Welch Allyn Welch Allyn 1 90 Welch Allyn Welch Allyn 90 WELCH ALLYN WELCH ALLYN...

Страница 22: ...16 Directions for Use Welch Allyn...

Страница 23: ...1 5 Welch Allyn 1 2 FlexiPortTM 3 4 5 1 Artery Index Marker Range 2 2 5 3 30 mm Hg 4 2 3 mm Hg 5 30 Welch Allyn 3 mm Hg 10 C 40 C 90 American National Standard ANSI AAMI SP10 2002 European Standard E...

Страница 24: ...Hg 300 250 200 150 100 50 60 0 mm Hg 3 0 mm Hg 6 0 mm Hg 0 1 mm Hg 3 1 mm Hg Welch Allyn Setra 2270 01 0 1 mm Hg Netech 200 2000IN 1 0 mm Hg Welch Allyn Welch Allyn Welch Allyn Welch Allyn Inc Welch A...

Страница 25: ...Welch Allyn 19 Welch Allyn Welch Allyn Welch Allyn Welch Allyn Welch Allyn Welch Allyn FlexiPortTM DuraShock Welch Allyn Inc FlexiPortTM 767 509...

Страница 26: ...20 Directions for Use Welch Allyn...

Страница 27: ...Welch Allyn 21 20000 Welch Allyn 20000 1 90 3 5 WelchAllyn 2 1 2 A 3 A 4 Welch Allyn Welch Allyn 3 1 2 11720 11730 11735 11470 11475 11710 3...

Страница 28: ...22 Directions for Use Welch Allyn...

Страница 29: ...gnetic interference and should not present problems to other equipment or be affected by other devices CAUTION Only authorized Welch Allyn Service Centers can perform service on this product CAUTION T...

Страница 30: ...24 Directions for Use Welch Allyn Attach Welch Allyn 3 5 V instrument heads of your choice to the handles See the directions for use for each instrument head 2 3 1 3 8 35 5 127 4 1 8 3 5 6 6...

Страница 31: ...utoclave the transformer or accessories WARNING Liquids can damage electronics inside the transformer Prevent liquids from spilling on or dripping into the transformer If liquids are spilled on or dri...

Страница 32: ...d manufacturer s declaration information is available at www welchallyn com GreenSeries777WallTransformer Operating Temperature limitation Humidity limitation Atmospheric pressure limitation Transport...

Страница 33: ...ed Alkaline Battery AA 1 5V IPX Not protected against the ingress of water MAX Latex Free Do not dispose of this product as unsorted municipal waste Prepare this product for reuse or separate collecti...

Страница 34: ...ety and temperature measurement accuracy use only accessories and supplies recommended or supplied by Welch Allyn Equipment is not suitable for use in the presence of a flammable anesthetic mixture wi...

Страница 35: ...y ingested medications age and gender A person s normal temperature tends to decrease with age The following table shows that ranges of normal also vary by site Do not directly compare measurements fr...

Страница 36: ...mp Plus thermometer as shown below Wall Holder Standard X Pulse Timer X Backlight X Security Electronic X Configurable On Screen Labeling for Identification of Instrument X Sold in U S Canada X X Sold...

Страница 37: ...gling of the probe cord with other equipment at the patient s bedside 1 Mount the instrument so that all other objects are no closer than 4 inches 10 25 cm from the sides and bottom of the unit 2 Moun...

Страница 38: ...show that the Oral Mode is selected Adult Axillary Mode This flashing icon and or words are displayed to show that the Adult Axillary Mode is selected Ped Axillary Mode This flashing icon and or words...

Страница 39: ...walking segments appear on the display indicating that measurement is in progress 7 The unit will beep three times when the final temperature is reached The measurement site temperature scale and pat...

Страница 40: ...ccurate readings 3 Verify that the probe tip is completely surrounded by axillary tissue and place the arm snugly at the patient s side Hold the patient s arm in this position and do not allow movemen...

Страница 41: ...separating the buttocks in place and hold the probe in place throughout the measurement process Rotating walking segments appear on the display indicating that measurement is in progress 6 The unit wi...

Страница 42: ...osition after the temperature is displayed 6 Press the Mode Selection button until the Monitor Mode indicator appears on the LCD 7 Hold the thermometer in place for the required duration 8 Record the...

Страница 43: ...ers from the probe cover storage compartment on the back of the unit 2 Remove the probe and the probe well 3 Press inward on the textured grips on each side of the assembly to remove the battery door...

Страница 44: ...can perform or approve all repairs on products under warranty Unauthorized repairs will void the warranty Qualified electronics personnel or at a Welch Allyn Service Center should repair out of warran...

Страница 45: ...risk Welch Allyn will not be responsible for loss associated with the use of any Welch Allyn product that 1 has had the serial number defaced 2 has been repaired by anyone other than an authorized Wel...

Страница 46: ...40 Directions for Use Welch Allyn...

Страница 47: ...in measurement error Connections 1 Connect one end of the coil tubing to the sphygmomanometer valve at the bottom of the unit 2 Align and press the FlexiPortTM adapter onto the cuff port 3 Attach the...

Страница 48: ...ing the cuff 1 Remove the bladder from the two piece cuff 2 Place the hook and loop fasteners in the closed position 3 Machine launder using gentle cycle warm water and mild detergent 4 Air dry comple...

Страница 49: ...warranties extend to the original purchaser and cannot be assigned or transferred to any third party This warranty shall not apply to any damage or product failure determined by Welch Allyn to have b...

Страница 50: ...44 Directions for Use Welch Allyn...

Страница 51: ...s Note It is important to note that while the throat illuminator section provides an excellent source of general illumination more then momentary direct skin contact should be avoided due to possible...

Страница 52: ...46 Directions for Use Welch Allyn Figure 1 Figure 2 Ophthalmoscopes No 11720 11730 No 11735 Ophthalmoscopes No 11470 11475 No 11710 Align Align Figure 3...

Страница 53: ......

Страница 54: ...llyn Medical Equipment Suzhou Co Ltd Floor 3 Building 9 No 8 Jinfeng Road Suzhou Science Technology Town Jiangsu Province China 215163 Telephone 86 512 6808 6078 Fax 86 512 6808 7806 8 9 3 215163 86 5...

Отзывы: