Button Functions
Setup
Assembling the Unit
Your SureTemp Plus thermometer is shipped to
you with three 1.5V AA batteries installed.
1. Install the oral/axillary probe (blue ejection
button) or optional rectal probe (red Ejection
button) by inserting the latching probe
connector on the end of the coiled cord into the
unit’s plug receptacle. Ensure that the
connector snaps into place. The LCD will
illuminate for approximately 7 seconds and will
then turn off.
2. Connect the appropriate removable probe well
to the thermometer. Note that the cut-out in the
probe well fits over the probe cord strain relief.
Ensure that the removable probe well “snaps”
into position and is completely seated to the
thermometer. Place the probe into the well.
3. Insert the box of Welch Allyn disposable probe
covers into the probe cover storage holder. To
open the box of probe covers, remove and
discard the tear-away corner. You will see the
ends of the probe covers in the box.
Mounting the Unit
Mount the wall-mount bracket in a location such as
an interior wall or wall divider so that the
thermometer remains at ambient room
temperature. Mounting the unit on an uninsulated
exterior wall, in an area directly hit by sunlight,
above heat radiators or powered equipment
(including computers and lights), or below air
conditioning outlets is not recommended.
Proper mounting of the unit will prevent tangling of
the probe cord with other equipment at the
patient’s bedside.
1. Mount the instrument so that all other objects
are no closer than 4 inches (10.25 cm) from
the sides and bottom of the unit.
2. Mount the wall mount a maximum of 48 inches
(122 cm) from the floor.
3. Install the wall mount to allow at least 12
inches 30.5 cm) clearance on top of the
instrument. This provides ease of removing
and installing the probe and probe well or
installing probe covers onto the probe.
Buttons
Procedure
C/F
Press and release the
C/F
(Celsius/Fahrenheit)
button to
toggle between temperature scales
any time a temperature is displayed
on the LCD.
Pulse Timer
(692 Only)
Press and release the
Pulse Timer
button to activate the timer. Audible
beeps sound at 0, 15, 30, 45, and
60 seconds. The timer turns off
automatically after 60 seconds. You
can stop the timer at any time by
pressing the
Pulse Timer
button or
by removing the probe from the
probe well.
Mode
Selection
Press and release the
Mode
Selection
button after you have
withdrawn the blue probe from the
probe well to change the current
measurement site to Oral, Adult
Axillary, or Pediatric Axillary. The
selected measurement site is
displayed by a flashing body site
icon. Note that no other
measurement sites are available
when using the red probe in the
Rectal Mode.
Recall
Press and release the
Recall
button when the display is blank to
recall the last completed predicted
temperature. The LCD will display
the recalled temperature for five
seconds along with the battery
icon, temperature scale, mode
selection, and probe-position icon
(if applicable).
ৎ㊩
ৎ㊩
ৎ㊩
ৎ㊩
Содержание GS 777
Страница 1: ...GS 777 GS 777 Integrated Wall System Directions for Use Material No 718848...
Страница 2: ......
Страница 6: ...vi Directions for Use Welch Allyn...
Страница 8: ...2 Directions for Use Welch Allyn Welch Allyn 3 5V 1 2 3 2 10 4 3 5V 1 2 3 4 1 8 3 5 6 6...
Страница 10: ...4 Directions for Use Welch Allyn...
Страница 19: ...Welch Allyn 13 5 SEC 30 ON OFF OFF 1 2 3 4 3 AA 1 5 3 AA 1 5 5 6 70 10 CaviCide 70 10 CaviCide AA 1 5...
Страница 21: ...Welch Allyn 15 1 Welch Allyn Welch Allyn 1 90 Welch Allyn Welch Allyn 90 WELCH ALLYN WELCH ALLYN...
Страница 22: ...16 Directions for Use Welch Allyn...
Страница 26: ...20 Directions for Use Welch Allyn...
Страница 28: ...22 Directions for Use Welch Allyn...
Страница 46: ...40 Directions for Use Welch Allyn...
Страница 50: ...44 Directions for Use Welch Allyn...
Страница 53: ......