Disassembly and Repair Section 4
Service Manual 488061 Rev. B Welch Allyn Exam Light III 3
7
. Light Pipe Connector Assembly Removal
With a 5/64 hex wrench, slightly back the hex
nut off the pipe connector assembly. Slide the
pipe connector assembly out. The pipe con-
nector assembly can only be removed from
the inside of the main housing
8. Transformer Removal
Remove the 4 phillips head screws.
NOTE: Note the wire orientation and wire
color on the illuminated off/on switch
L
Light
Blue
Blue
Black
Brown
. The
wires must be removed from the illuminated
switch before the illuminated switch can be
removed.
The photo shows the wiring as wired for
domestic 120V. See the wiring diagram in
Appendix C for the wiring in your local area.