Functional verification
•
In WAST, double click the device and go to
verify and calibrate
•
To verify and calibrate the complete device,
select the device and click
perform all
•
Alternatively you may perform each module
separately but device calibration date will not
be changed nor a calibration certificate
generated.
•
Follow the instructions on screen to perform
all tests and calibrations
Содержание CSM 7100
Страница 1: ...Connex Spot Monitor CSM Service and Repair training ...
Страница 27: ...Lab Use Menu Advanced menu ...
Страница 28: ...Recommended service intervals ...
Страница 39: ...Disassembly Procedures Backhousing ...
Страница 40: ...Disassembly Procedures Backhousing SureTemp and Pro6000 dock procedures are the same ...
Страница 41: ...Disassembly Procedures Speaker Disconnect cable and slide speaker out ...
Страница 44: ...Disassembly SpO2 board Procedures differ per SpO2 type Nonin Covedien Nellcor Massimo ...
Страница 46: ...Disassemble LCD Pull out the LCD flex cable friction Turn assembly over and lift LCD as indicated ...
Страница 47: ...Disassemble LCD Remove the gaf tape 1 Carefully lift the bale to unlock the flex cable 2 Remove the LCD 1 2 ...
Страница 48: ...Disassemble Pump Assembly Disconnect tubing manifold and connector ...
Страница 49: ...Disassemble Pump Assembly Remove 2 screw that hold the pump assembly Pump assembly can now be removed ...
Страница 50: ...Disassemble Bezel The rubber bezel is locked with the 6 bosses on the housing ...
Страница 54: ...Reassembly torque value table ...
Страница 63: ...Functional verification The WAST help file displays all required tools and connections ...
Страница 65: ...Hands on Functional verification Let s make sure it s working as it should Lab Functional verification ...
Страница 67: ...Configuration tool http config welchallyn com ...
Страница 69: ......
Страница 70: ......