Directions for Use
Chapter 4 Performing Spirometry Tests
37
Connecting the Spirometer Components
1.
Verify that the sensor and pressure tubing are clean and undamaged. Look for signs
of deterioration, including but not limited to cracks, cuts, discoloration, or oxidation.
If any part exhibits any of these symptoms, replace it. See
2.
Attach a flow transducer to the pressure tubing. See
.
3.
Attach a sensor to the other end of the pressure tubing. See
.
4.
Connect the sensor to the electrocardiograph’s spirometry port. Hand-tighten the
sensor connectors. Do not overtighten the connectors, or they may become stripped.
See
.
The CP 200 software automatically activates the spirometry functions throughout the
software.
Figure 26. Connecting the Sensor to the Spirometry Port
WARNING
To prevent the spread of infection, use a new flow transducer for
each patient. Use rubber gloves when replacing used flow transducers, and wash
hands after touching them. Discard flow transducers after a single patient use.
Note
Bacteria filters are unnecessary.
Figure 24. Attaching a Flow Transducer
to the Pressure Tubing
Figure 25. Attaching the Sensor to the
Pressure Tubing
Содержание CP 200
Страница 1: ...CP 200 Spirometry Option Directions for Use ...
Страница 34: ...30 Chapter 3 Calibrating the Spirometer Welch Allyn CP 200 Spirometry Option ...
Страница 52: ...48 Chapter 4 Performing Spirometry Tests Welch Allyn CP 200 Spirometry Option ...
Страница 53: ...49 5 Troubleshooting Problem Solving Suggestions 50 Limited Warranty 51 Service Policy 51 ...
Страница 56: ...52 Chapter 5 Troubleshooting Welch Allyn CP 200 Spirometry Option ...
Страница 80: ...76 Appendix D Predictive Norms etc Welch Allyn CP 200 Spirometry Option ...
Страница 91: ......