Note
New batteries are approximately 30 percent charged. Therefore, connect the
battery to AC power immediately after inserting a new a battery.
WARNING
Risk of fire, explosion, and burns. Do not short-circuit, crush,
incinerate, or disassemble the battery pack. Never dispose of batteries in refuse
containers. Always recycle batteries according to local regulations.
Clean the monitor
WARNING
Electric shock hazard. Before cleaning the monitor, disconnect the AC
power cord from the power outlet and the monitor.
WARNING
Electric shock hazard. DO NOT autoclave the monitor or accessories.
The monitor and the accessories are not heat-resistant.
WARNING
Liquids can damage electronics inside the monitor. Prevent liquids
from spilling on the monitor.
If liquids are spilled on the monitor:
1. Power down the monitor.
2. Disconnect the power plug.
3. Remove battery pack from the monitor.
4. Dry off excess liquid from the monitor.
Note
If liquids possibly entered the monitor, remove the monitor from use
until it has been properly dried, inspected, and tested by qualified
service personnel.
5. Reinstall battery pack.
6. Power on the monitor and verify monitor functions normally before using it.
If liquids enter the printer housing:
1. Power down the monitor.
2. Disconnect the power plug.
3. Remove battery pack from the monitor.
4. Remove and discard the paper roll.
5. Clean and dry the inside of the printer housing.
Note
The printer housing has a drain tube that directs liquids down and
out the bottom of the device. If liquids possibly entered other
openings in the monitor, remove the monitor from service until it has
been properly dried, inspected, and tested by qualified service
personnel.
6. Install a new roll of paper.
7. Power on the monitor and verify that the monitor functions normally before using it.
Clean on a routine basis according to your facility's protocols and standards or local regulations. If
the monitor is on, lock the display and disconnect the AC power cord.
The following agents are compatible with the monitor:
•
CaviWipes
™
•
Sani-Cloth
®
Plus
•
70 percent isopropyl alcohol
•
10 percent chlorine bleach solution
90
Maintenance and service
Welch Allyn Vital Signs Monitor 6000 Series™
Содержание Connex Vital Signs Monitor 6000 Series
Страница 1: ...Welch Allyn Vital Signs Monitor 6000 Series Directions for use ...
Страница 6: ...vi Contents Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 8: ...2 Introduction Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 14: ...8 Screen elements Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 22: ...16 Controls indicators and connectors Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 30: ...24 Setup Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 40: ...34 Navigation Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 44: ...38 Profiles Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 64: ...58 Alarms Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 92: ...86 Patient monitoring Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 98: ...92 Maintenance and service Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 106: ...100 Specifications Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 134: ...128 Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series ...
Страница 143: ......