AHA
RL - Green
RA - White
LA - Black
V - Brown
LL - Red
Impedance respiration
In some patients, impedance respiration detection may be inadequate using the standard
ECG electrode placement. In these cases, change the LL and RA electrode placement to
the mid-axillary line on each side of the chest as shown in the illustrations.
Note
Respiration can only be acquired from Lead II.
IEC
N - Black
R - Red
L - Yellow
C - White
F - Green
AHA
RL - Green
RA - White
LA - Black
V - Brown
LL - Red
Directions for use
ECG acquisition 21