21
PREVENTIVE MAINTENANCE
Preventive Maintenance Schedule
Maintenance to be
Performed
RECOMMENDED
FREQUENCY
NOTES
Device Cleaning
As Needed
Refer to the device cleaning portion of this section.
Preventive Maintenance
Procedure
12 Months
Refer to procedure defined below.
Battery Replacement
24 Months
Replacement period varies based on frequency and
duration of use; refer to the Battery Health value
displayed in the device special functions for specific
information regarding battery capacity.
WARNING: Preventive maintenance is to be performed by Welch Allyn authorized service
personnel only.
Preventive Maintenance / Conformance Testing
Recommended Equipment
Multi-Meter
ECG Simulator
AM12 Patient Input Module (9293-048-5X)
WAM Patient Input Module (30012-019-5X)
WAM/AM12 Lead Wire Set (9293-046-70)
AM15E Patient Input Module (9293-063-50)
LEAD SET AM15 (9293-046-80)
Patient Cable Snap Adapter Set (9281-002-50)
10 Lead Shorting Block (or equivalent)
Lead Test Failure Box (or equivalent)
PC with ELI-Link v3.10 or later
USB Cable Type A to B (6400-012)
PC with NIC and ELI-Link v4.5 or later
Wireless Router 802.11 (a, b, g, n)
FAT 32 USB Memory Device
Electrical Safety Analyzer
Ruler (mm)
Smart Thermal Paper
Preventive Maintenance Procedure
Print the device configuration (attach to the Preventative Maintenance Data Record / PMR).
Remove the unit cover per Unit Disassembly section of this manual, and perform a visual inspection of the
following items: