15
NOTE: THIS FIRMWARE UPDATE PROCEDURE REQUIRES A LAPTOP/PC WITH WINDOWS 10 AND A USB
TYPE A TO B CABLE.
Firmware Upgrade Process
1.
Obtain the required firmware file (exe file) from Welch Allyn.
2.
Record the serial number of the Scale.
3.
With the Scale powered OFF, connect the scale to the laptop/PC with a USB Type A to B cable. Check
that the Scale is being detected by the laptop/PC in Device Manager as
“USB Serial Device” (Refer to
below image). (Note: If this is the first time the Scale is plugged into the laptop/PC, the driver will be
automatically installed.). The port number may or may not be COM9.
4.
Double click on the exe file. A dialog box will appear (Refer to below image)
5.
Click on the “Start” button shown in dialog box. Select the COM PORT to which the Scale is
connected. (Refer to image below as an example).
6.
Click “Ok”. The following dialog box will appear.
Содержание 5002 Series
Страница 6: ...6 ...
Страница 7: ...7 ...
Страница 9: ...9 ...
Страница 30: ...30 1 Remove handrail screws 6 and handrail 2 Remove mast bolts and washers 2 and mast ...
Страница 34: ...34 iii Remove printer from the door 7 Remove battery door screws 2 and battery door 8 Remove batteries ...
Страница 51: ...51 Base Disassembly overview 6002 Base 5702 Base 5002 Base 6702 Base 6702W Base 6702SP Base ...
Страница 56: ...56 11 Remove load mounts 4 12 Remove load cell nuts 8 and load cells 4 13 Remove overload protection plates 6 ...
Страница 85: ...85 5122 Disassembly Mast Removal and Disassembly 1 Remove the slider mast cover mounting screws 2 and cover ...
Страница 89: ...89 8 Remove batteries 6 9 Remove battery holder assembly mounting screws 4 and battery holder assembly ...
Страница 91: ...91 10 Remove power jack mounting screws 2 and power jack ...
Страница 112: ...112 Full Assembly Exploded Views 5002 ...
Страница 115: ...115 5122 ...
Страница 118: ...118 5702 ...
Страница 121: ...121 6002 ...
Страница 124: ...124 6702 ...
Страница 127: ...127 6702W ...
Страница 130: ...130 6702SP ...