88
Troubleshooting
EN
* These alarms are only activated 2 minutes after
ventilation starts.
8.2.5
Technical alarms
Caution!
If faults occur which cannot be eliminated at once, contact the manufacturer,
Weinmann, or your specialist dealer immediately to have the device repaired. Do
not continue operating the device in order to prevent even greater damage.
Apnea
Low priority
No spontaneous
breathing for at least 3
breaths
Have the settings checked by
the doctor supervising
treatment
Apnea
(Only in MPVv and
MPVp modes)
High priority
No spontaneous
breathing within set
time
Have the settings checked by
the doctor supervising
treatment
Leakage
(Only available in
dual leakage
ventilation system)
Medium priority
Leakage
Find and eliminate leak,
replace patient circuit if
necessary
Display
Alarm
Cause of fault
Remedy
Display
Alarm
Cause of fault
Remedy
Battery capacity
critical
Medium priority
Battery discharged
(under 25 % capacity
remaining)
Restore power supply and
charge battery. In the event of
an extended power outage,
keep an alternative ventilation
option to hand or use a
replaceable battery.
Battery capacity
highly critical
High priority
Battery discharged
(under 10 % capacity
remaining)
Restore power supply and
charge battery. In the event of
a power outage, keep an
alternative ventilation option
to hand or use a replaceable
battery.