80
EN
Subject to design modifications.
1 hPa = 1 mbar
Precision of dynamic
pressure (short-term
precision) with 15
breaths/min to EN ISO
17510-1:2009 at
4 hPa
8 hPa
11 hPa
15 hPa
18 hPa
Δ
p = 0,2 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,2 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Precision of dynamic
pressure (short-term
precision) with 20
breaths/min to EN ISO
17510-1:2009 at
4 hPa
8 hPa
11 hPa
15 hPa
18 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,4 hPa
Δ
p = 0,4 hPa
Δ
p = 0,5 hPa
Δ
p = 0,3 hPa
Δ
p = 0,3 hPa
Δ
p = 0,4 hPa
Δ
p = 0,4 hPa
Δ
p = 0,5 hPa
Precision of static pressure
(long-term precision) to
EN ISO 17510-1:2009
Δ
p = 0,04 hPa
Constant pressure level
measured as per
prEN ISO 17510
with 6
mbar:
Δ
p = 0.3 mbar
with 12 mbar:
Δ
p = 0.4 mbar
with 18 mbar:
Δ
p = 0.6 mbar
Fine filter separation level
to 1
μ
m
to 0.3
μ
m
≥
99,5%
≥
85%
Fine filter service life
≥
250 hours with normal room air
from device no. 4000 and
software version 04/02
upwards:
Trigger threshold
Can be set to 5 levels each for inhalation and exhalation