22
EN OXYWAY
WM
6658
1a
11
/20
1
7
5 Hygienic preparation
5.3 Hygienic preparation of pressure
reducers
Requirement
The pressure reducer is removed
(see “4.2 Removing the pressure
.
1. Carry out hygienic preparation of the pressure reducer as
specified in the following table:
2. Install the pressure reducer
(see “4.1 Installing the pressure
.
3. Perform a function check
(see “6.2 Performing a function
.
Result
The pressure reducer is hygienically prepared.
Risk of injury if disinfectants containing alcohol or grease are
used for wipe disinfection!
Disinfectants containing alcohol or grease react with oxygen to
create a flammable mixture and can result in explosions and injury
to the user and patient.
When wiping the pressure reducer with disinfectant, do not
use disinfectants containing alcohol or grease.
Risk of injury due to ingress of liquids!
Liquids entering the pressure reducer may cause explosions and
therefore injure the patient.
Do not immerse the pressure reducer in liquids.
Part concerned Cleaning
Disinfection
Thermal
disinfection
Sterilization
Pressure reducer
Using a lint-free, dry
or moistened cloth:
Use clean water.
Wipe disinfection
(Recommendation:
terralin
®
protect
)
Not permitted
Not permitted
Содержание OXYWAY