
WM
6807
1
0
5
/2
021
MEDUMAT Standard
2
EN
225
11 Maintenance
11.4 Maintaining the reusable breathing
circuit
Requirement
The reusable breathing circuit has been disassembled
(see "8.4 Disassembly of the reusable breathing circuit",
page 176)
1. Check all parts of the reusable breathing circuit for external
damage and complete labeling.
If necessary: Replace damaged or incorrectly labeled parts.
2. Replace the PEEP control diaphragm and check valve
diaphragm (maintenance kit WM 15779).
3. Assemble the reusable breathing circuit
reusable breathing circuit", page 190)
.
4. Punch out the date at which the next maintenance is due
on the service label (maintenance kit WM 15779).
5. Attach the service label to the end of the ventilation hose
which is closest to the device.
6. Perform a function check
(see "9.3 Performing a function
Result
The reusable breathing circuit has been maintained and is ready
for use.
Содержание MEDUMAT Standard2
Страница 1: ...MEDUMAT Standard2 Ventilator Instructions for Use for Devices from Software Version 5 1...
Страница 110: ...110 EN MEDUMAT Standard2 WM 68071 05 2021 5 User menu 5 3 Settings in the user menu 5 3 1 Mode 5 2 Submenu Mode...
Страница 269: ......
Страница 270: ......
Страница 271: ......