8 Hygienic reprocessing
8.9 Preparing parts for reuse
Requirement
The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage due to use (e.g., tension cracks and
cable breakage).
2. Replace damaged parts.
3. Assemble the reusable breathing circuit
reusable breathing circuit", page 190)
4. Mount the accessories.
5. Reconnect the power supply
(see "4.2 Connecting to a power
6. Perform a function check
(see "9 Function check", page 200)
7. Place parts in storage in accordance with the storage
conditions
(see "14.1 Technical data", page 231)
Result
The parts are ready for use again.
8.10 Assembling reusable breathing circuit
The images in this subchapter show all the possible components
of the reusable breathing circuit. Depending on the system type,
your reusable breathing circuit may not include certain
components
(see "3.5.2 Reusable breathing circuit and disposable
.
Requirement
The reusable breathing circuit is disassembled.