26
Repair information and repair instructions
6. Slide the contact rocker into its resting position.
7. Check to see that the contact pin is directly above
the capacitor contacts.
If necessary, make some fine settings using the
black screw on the front of the carriage.
8. Flip down the contact rocker. The contact pin must
now be touching the capacitor contacts on the
printed circuit board.
The capacitor is then completely discharged. The
discharging process lasts approx. 4 seconds. It is
completed when
-0.00
is displayed.
To be certain that the capacitor is discharged, carry
out the following tests:
9. Press the red button on the front of the device.
– The capacitor is very slightly charged. The
display will now show for example
0.07
.
– Upon releasing the button the display must
slowly
return to
-0.00
.
Danger! Risk of electric shock!
If the display returns immediately to
-0.00
then the
contact pin is no longer in contact with the capacitor
contact on the printed circuit board. The capacitor has in
this case not been discharged. Adjust the position of the
contact pin and check it again.
Completion
1. Pull the contact rocker back.
2. Slide the locking element back to release the
lifting aid.
3. Lower the upper part of the encasing.
4. Rotate the lifting aid to release the suction pad
from the lid.
5. Open the protective cover of the device and
remove the MEDUCORE Easy.
6. Switch the discharging device via the rocker
switch off.
7. Close the device (8.6, page 28).