Description
11
EN
•
permits respiration-controlled oxygen inhalation
in Demand mode.
3.2 Owner/operator and user qualification
As an owner/operator or user, you must be familiar
with the operation of this medical device. Observe
the legal requirements for operation and use (in Ger-
many, the regulations governing owner/operators of
medical devices apply in particular). Basic recom-
mendation: get a person authorized by WEINMANN
Emergency to provide you with proper instruction
about the handling, use and operation of this medi-
cal device.
3.3 Ventilation function
MEDUMAT Easy operates within a pressure range of
2.7 to 6 bar and at a flow rate of not less than
70 l/min O
2
. It has a built-in power pack.
It uses high-pressure, medicinal-grade oxygen. An
external pressure reducer brings this down to the re-
quired operating pressure. The oxygen supply is fed
in at input valve.
The ventilation settings are continuously variable.
These settings (frequency and volume per minute
are coupled) and the inspiration/expiration ratio of
1:1.67 are regulated by internal electronic control
mechanisms.
The gas for inspiration flows along the hose and
through the patient valve and either the mask or
tube into the patient’s airways. The patient valve is
16
15
13
14
14
12
12
11
10
10
9
11
16
7
5
30
3
MV (l/min)
Freq.(min
-1
)
Demand-
flow
10
0
Содержание Emergency MEDUMAT Easy
Страница 2: ...English 3 Espa ol 70 Portugu s 146...
Страница 5: ...Overview 5 EN B 10 14 13 11 12 15 C 16 17 19 20 18 21 or...
Страница 69: ......