SLS Baseline 8900 Setup and Calibration Manual 5-1-GL-GL-SUL-00028
Page 17
SKYCHROM SEQUENCE
In order to operate, the Baseline 8900 must also have a sequence. A very basic sequence of receiving
the sample from sample port 001 is all that is required for operation and a loop of this sequence will be
used for operation. The export button must be clicked in order for the results of this sequence to export
to WellWizard. Chromatograms may also be saved under certain circumstances, or all chromatograms
may be saved using the bullet menu on the right of the sequence window as seen in Figure 10. The CHM
button will allow a title to be applied to those saved chromatograms and will be seen in the file window
at the bottom of the window.
Figure 10: Skychrom Sequence
All sequences must be saved in order to function. In order to do this, the sequence window must remain
open in Skychrom and the user must navigate to the “Sequence” at the top menu. Select “Save as” and
save the sequence. Figure 11 demonstrates this process.
Содержание SLS Baseline 8900
Страница 10: ...SLS Baseline 8900 Setup and Calibration Manual 5 1 GL GL SUL 00028 Page 9 Figure 2 Pneumatic Xpress Setup ...
Страница 17: ...Page 16 5 1 GL GL SUL 00028 SLS Baseline 8900 Setup and Calibration Manual Figure 9 Method Spline ...
Страница 19: ...Page 18 5 1 GL GL SUL 00028 SLS Baseline 8900 Setup and Calibration Manual Figure 11 Sequence Save ...
Страница 24: ...SLS Baseline 8900 Setup and Calibration Manual 5 1 GL GL SUL 00028 Page 23 Figure 16 Skychrom Export ...