background image

ELITE 600 V1.0 

A. Important Notice

 

 

Before setting up and operating the instruments, please carefully read these instructions to 

get familiarized with the installation and operation process. Instructions should be read by 

experienced individuals before operating the instruments.     

Any improper usage of the instruments may cause damage. Please refer to the safety 

notice included with this equipment. 

The instruments shall not be modified or altered in any way. Any modification or alteration 

will void the warranty, void the regulatory certifications and create potential safety hazard. 

Wealtec is not responsible for any injury or damage caused by using the instruments for any 

non-intended purpose or injury as a result of modification of the instruments by any person who 

is not authorized by Wealtec Corp.   

 

A-1. Warranty 

 

ELITE 600 are warranted to be free from defects in materials or workmanship for a period 

of one year from the original invoice date, under normal usage. Any defects occurring during 

warranty period, Wealtec Corp. will repair or replace defective products or parts without charge 

unless the defects arise from conditions outlined below. The defects described below are 

specially excluded from Wealtec warranty policy.   

1.  Improper operation of the instrument. 

2.  Repair or modification by any person who is not authorized by Wealtec Corp.   

3.  Damage caused by any (in)-direct accident, neglect or misuse. 

4.  Damage caused by disaster.   

5.  Damage caused by any improper solvents or samples 

 

A-2. Technical and Service Contact 

 

Most of the operation details are described in this instruction manual to assist and guide 

operator for an appropriate solution. For any other technical/ service questions, please contact 

your local representative or contact Wealtec international technical/ service specialist by E-mail: 

[email protected]

 

  1

Содержание ELITE 600

Страница 1: ...ELITE 600 Power Supply Installation and Operation Manual Version 1 0 Item 02013 This instrument is intended for laboratory use only ...

Страница 2: ...ct Description B 2 1 ELITE 600 Power Supply Hardware Overview C Installation of ELITE 600 Power Supply 5 C 1 Package List C 2 Setup and Operation of the ELITE 600 Power Supply C 2 1 Unpacking C 2 2 Setup and operation D Troubleshooting 8 D 1 Basic troubleshooting D 2 Error codes D 3 Replacing the fuses E Care and Maintenance 11 F Order Information 12 i ...

Страница 3: ...nty ELITE 600 are warranted to be free from defects in materials or workmanship for a period of one year from the original invoice date under normal usage Any defects occurring during warranty period Wealtec Corp will repair or replace defective products or parts without charge unless the defects arise from conditions outlined below The defects described below are specially excluded from Wealtec w...

Страница 4: ...lways connect the power supply using the AC power cord provided with the power supply or a similar 3 prong grounded AC outlet Use other than Wealtec s electrical cable with plugs or banana jacks is done at the user s own risk when connecting the external instrument to the power supply s high voltage output jacks While doing so always grasp the plug by the molded support at the rear of the plug Do ...

Страница 5: ...ent modes Power output is controlled by setting a maximum value for voltage up to 600 V or current up to 500 mA The maximum output power at any time is 90 Watts The Run duration can be continuous or programmed to a maximum duration of 999 minutes The ELITE 600 automatically crosses over or switches the limiting parameter according to the programmed limits as the resistance of the electrical load c...

Страница 6: ...value of the selected parameter 5 RUN Starts and Stops the output from the Power Supply The front and rear panel buttons as shown above are used to set up and run the ELITE 600 power supply The audible alarm design allows a beep sound for each and every button pressed on the front membrane keypad A long beep alarm allows user to identify the run is complete or if an error occurs It will continue e...

Страница 7: ... was in transit If any part appears damaged contact the carrier immediately Be sure to keep all packing material for damage claims or to use should it become necessary to return the unit C 2 2 Setup and Operation Caution Use only one hand when making or breaking a connection to avoid making a complete circuit across your chest Also while connecting or disconnecting the leads do not allow any part ...

Страница 8: ... to the maximum output of the power supply This is the normal condition or setting for simple electrophoretic separations Optional If there is another upper limit desired for this non constant variable use the up and down buttons to set it to the desired value 7 Set the timer Optional Press the MODE button to move the green indicator LED light to under clock symbol Continuous run Leave the timer a...

Страница 9: ...un time in minutes Press both the UP and DOWN buttons at the same time The display will briefly show PF0 indicating that the PFD mode is disabled Press both the UP and DOWN buttons at the same time again The display will show PF1 indicating that the PFD mode is enabled 8 Start the run Press the RUN button The green LED light above the RUN button indicates that voltage and current are being applied...

Страница 10: ...heck if power switch malfunction 4 Contact your local Wealtec representative Fan is spinning LED Display do not light up Front panel power on indicator off 1 Check main power source 2 Check if any errors on the front panel LED display 3 Contact your local Wealtec representative Fan is spinning LED Display do not light up Front panel power on indicator on 1 Contact your local Wealtec representative...

Страница 11: ...as detected at some point after a successful start Buffer is leaking from the cells There is a loose electrical connection in the circuit 1 Check and correct all potential resistance problems 2 Check buffer volumes 3 Check all voltage leads 4 Check the power supply adapters fitting If used 5 Press any button to clear the error code E4 Power Failure during a Continuous run no timer used 1 Restart t...

Страница 12: ...code Press any button except the RUN button to clear the error message D 3 Replacing the fuses 1 Unplug the power cord from the power supply 2 Locate the fuse compartment on the rear panel 3 Use a screwdriver to pry out the fuse compartment There are two fuses in the compartment Check and replace any blown fuses with a fuse of the same size and ratings 4 FUSES 2 250 V 3 15 A 5 x 20 mm fast blow 10...

Страница 13: ...he power supply and allow to dry completely Please contact your local representative or Wealtec international technical service specialist by E mail support wealtec com Wealtec Corp offers complete technical support for all of our products If you have any questions about how to use this product or would like to arrange to repair it please contact your local Wealtec Corp representative or internati...

Страница 14: ...TE 300 Power Supply 120 V 50 60 Hz 1001004 ELITE 300 Power Supply 230 V 50 60 Hz 1001005 ELITE 300 Plus Power Supply 120 V 50 60 Hz 1001006 ELITE 300 Plus Power Supply 230 V 50 60 Hz 1001011 ELITE 600 Power Supply 100 240V 50 60 Hz 1001007 Mini ELITE Power Supply 90 264 V 50 60 Hz Standard package Complete power supply unit an instruction manual power cable and a warranty card ...

Страница 15: ...02013 Wealtec Corp 1885 Meadowvale Way Sparks NV 89431 Tel 1 775 351 2066 Fax 1 775 351 2077 http www wealtec com E mail support wealtec com Printed in Taiwan ...

Отзывы: