PREPARATION AND CAUTION BEFORE
MEASUREMENT
1.
Before measurement, warm up for at least 30 seconds, after connecting the
thermocouple to the connector of the Temperature Converter.
2.
If the instrument is used near noise generating equipment, be aware that the
Voltage output may become unstable or indicate large errors.
3.
Check TC-253A battery. The multimeter should read > 7.1 volts in the OFF
position.
TEMPERATURE MEASUREMENTS
1.
Warning!
do not measure the temperature of any surface that the potential
exceeds 60 Vdc or 24 Vrms.
2.
Connect the plug of the thermocouple to the connector of the thermometer.
3.
Select the ºC/ºF function desired.
4.
Insert the “Output Plugs (V and COM)” to the input terminals of Multimeter. Set
the Multimeter to “DC mV or V” range.
5.
Place the sensing point of thermocouple on the surface or environment to be
measured.
6.
Wait until the reading stabilizes.
GENERAL SPECIFICATIONS
This Temperature Converter conforms to the temperature/voltage tables of the National
Bureau of Standards and to the IEC 584 Standards for K-type.
1.
Temperature Scale: Celsius or Fahrenheit user selectable.
2.
Input: Single K-type thermocouple.
3.
Output to Meter: 1mVdc per ºC or ºF.
4.
Measurement Range: -50ºC to 1000ºC – (-58ºF to 1832ºF).
5.
Low Battery Indication: if the reading is less than 7.1V.
6.
Temperature Coefficient: 0.15 x (specified accuracy) / ºC, (<18ºC or >28ºC).
7.
Power Requirements: Standard 9V battery, NADA 1604, JIS 006P, or IEC6F22
Size.
8.
Battery Life: Alkaline 300 hours.
9.
Dimensions: 122mm(L) x 46mm(W) x 30mm(D).
10.
Weight (including battery): 114 grams.
11.
Accessories: K-type bead thermocouple, battery (installed), operators manual.