– 5 –

VOLTAGE MEASUREMENT

1. Turn off power to the device under test and discharge all capacitors.
2. Set the Function/Range switch to the desired voltage type (AC or DC)

range.

3. Plug the black test lead into the COM jack on the DMM and connect

the test lead point to a grounded point (the reference point  for mea-
surement of voltage).

4. Plug the red test lead into the V jack on the DMM and connect the test

lead point to the circuit where a voltage measurement is required.
Voltage is always measured in parallel across a test point.

5. Turn on power to the circuit/device to be measured and make the volt-

age measurement. Reduce the range setting if set too high until a sat-
isfactory reading is obtained.

6. After completing the measurement, turn off power to the circuit /

device under test, discharge all capacitors and disconnect the DMM
test leads.

CURRENT MEASUREMENT

1. Turn off power to the device under test and discharge all capacitors.
2. Set the Function/Range switch to the desired current range.
3. Plug the black test lead into the COM jack on the DMM and connect

the test return branch for the current to be measured.

4. Plug the red test lead into the 200mA current jack of the DM2.

Current measurements are always made with the meter in series with
the test branch. The circuit must be broken and the meter put in series
by placing the red test lead point to the test node for current.

5. Turn on power to the circuit/device under test and make the current

measurement.  Reduce the range setting if set too high until a satisfac-
tory reading is obtained.

6. After completing the measurement, turn off power to the circuit/device

under test, discharge all capacitors, disconnect the DMM test leads
and re-connect the circuit branch.

Note: 

The current jack has a protective fuse installed to protect the

meter from damage and the user from harm if a current larger
than specified is applied at the input.  If the quick acting fuse
blows, replace with a like fuse.

DM2.Man.08.00  11/29/00  8:28 PM  Page 5

Содержание Meterman DM2

Страница 1: ...TM Digital Multimeter Multi Tester Operator s Manual Model DM2 Digital Multimeter Manual de Instrucciones...

Страница 2: ...t limited to implied warranties of merchantability and fitness for a particular purpose are limited in duration to the above stated on 1 year period Wavetek Meterman shall not be liable for loss of us...

Страница 3: ...on the instrument itself Exercise extreme caution when measuring voltage 20 V current 10mA AC power line with inductive loads AC power line during electrical storms cur rent when the fuse blows in a c...

Страница 4: ...nd Range desired 2 COM Input Terminal Ground input connector 500V DC AC Peak Maximum voltage between COM and earth ground 3 V Input Terminal Positive input connector for voltage resistance and diode t...

Страница 5: ...splayed If overload condition still exists in the highest range remove test leads from the measurement setup as it is beyond the range of the meter 1000 200 20 2 200 m 20 u 200 2m 20m m 200 k 2000 500...

Страница 6: ...wer to the device under test and discharge all capacitors 2 Set the Function Range switch to the desired current range 3 Plug the black test lead into the COM jack on the DMM and connect the test retu...

Страница 7: ...o be subtracted from resis tance reading Open circuits will be displayed as an overload condition DIODE TEST Diode Test checks the forward bias of diode function 1 Turn off power to the device under t...

Страница 8: ...Relative Humidity Storage Temp 20 C to 60 C 0 to 80 R H with bat tery removed from meter Accuracy Stated accuracy at 23 C 5 C 75 R H Power Stated 9 volt battery NEDA1604 JIS 006P IBC 6F22 Battery Life...

Страница 9: ...t 1 0 0 6mA Test Voltage 3 2VDC max MAINTENANCE WARNING To prevent electrical shock hazard turn off the multimeter and any device or circuit under test and disconnect the test leads before removing th...

Страница 10: ...placement fuse of the proper rating as shown above Cleaning the Meter Use only mild detergent and warm water to clean the meter Do not use aromatic hydrocarbons or chlorinated solvents REPAIR Read the...

Страница 11: ...ime on out of warranty repairs please phone in advance for service charge rates The multimeter should be shipped with transportation charges prepaid to one of the following addresses or to a service c...

Страница 12: ...TM Digital Multimeter Multi Tester Manual de Instrucciones Model DM2 Digital Multimeter...

Страница 13: ...se har responsible de p r didas de uso del mult metro ni de ning n otro da o accidental o con secuencial gastos o p rdidas econ micas en ninguna reclamaci n a que pudiera haber lugar por dichos da os...

Страница 14: ...circuitos expuestos ni partes met licas Mantenga su cuerpo aislado de tierra No utilice el instrumento en ambientes potencialmente explosivos ACCESORIOS OPCIONALES VC12 Estuche de vinilo acolchado RTL...

Страница 15: ...ediante de un 1 o 1 en el visualizador Para eliminar la causa de sobre carga seleccione un escala m s alta Si ya est en la escala m s alta interumpa la medida La indicac on de sobrecarga es normal en...

Страница 16: ...iempe el dispoitivo o circuito sometido a prueba y descarge todos los condensadores 2 Ponge el selector de funcion a la escala deseada de corriente 3 Conecte la punta de prueba negra a la entrada de C...

Страница 17: ...OL 3 Conecte la punta de prueba roja a la entrada V la negra a la entrada COM 4 Aplique la punta de prueba roja al anodo del diode y la negra al cato do El visualizador indica la ca da de tension dire...

Страница 18: ...R Temperatura de almacenamiento 0 20 a 60 C 0 a 80 R H sin pila Excactitud 23 C 5 C 75 R H Alimentaci n Pila normal de 9V NEDA 1604 JIS 006P IEC 6F22 Duraci n de la pila tipica 200 horas alcalina 150...

Страница 19: ...ecisi n 1 2 de la lectura 10 digitos Lastre de tensi n 350V maximo Protecci n sobrecarga 8A fusible de 250V Resistencia Escalas 200 2k 20k 200k 2M Precisi n Escala de 200 1 8 de la lec tura 4 digitos...

Страница 20: ...le abierto por otro nuevo de mismo valor Cambio de la Pila 1 Quite los dos tornillos y separe la tapa posterior del medidor 2 Extraiga la pila y cambiela por otra equivalente de tipo NEDA 1604 9 Vuela...

Страница 21: ...mero de tel fono prueba de com pra para reparaciones en garant a una breve descripci n del prob lema o el servicio requerido y en caso de reparaciones fuera de garant a si desea presupuesto previo Por...

Страница 22: ...g from Wavetek Meterman Specifications subject to change Wavetek is a trademark of Wavetek Wandel Golterman Wavetek Meterman 2000 U S Service Center Wavetek Meterman 1420 75th Street SW Everett WA 982...

Отзывы: