Wavetek 15XP Скачать руководство пользователя страница 12

10

Specifications

Display:
5XP and 15XP: 

3 ½ digit liquid crystal

display (LCD) with a maximum reading of
1999.

35XP: 

3 ¾ digit liquid crystal display

(LCD) with a maximum reading of 3999.

Polarity: Automatic, positive implied,
negative polarity indication.
Overrange: 

0o

 or 

-0o

 is displayed.

Low battery indication: The 

M

symbol is displayed when the battery
voltage drops below the operating level.
Operating environment: 0 °C to 50 °C
at <70 % R.H.
Power: Single standard 9 V battery,
NEDA 1604, JIS 006P, IEC 6F22.
Dimensions: 155 mm (H) ×72 mm (W)
×32 mm (D)
Weight: Approximately 210 g including
battery.
Overload protection: 1000 V dc or
750 V ac

Accessories: 

One pair test leads TL245,

9 V battery (installed),holster, magnet
strap, Users Manual.
Altitude: 6561.7 Feet (2000 M)

 

Safety: Conforms to EN61010-1, Rev-2;
CAT I 1000V, CAT II 600V, CAT III 300V,
class 2 and pollution deg.2;
CSA 22.2 -1010-1.
EMC: Conforms to EN61326-1. This
product complies with requirements of the
following European Community Directives:
89/ 336/ EEC (Electromagnetic
Compatibility) and 73/ 23/ EEC (Low
Voltage) as amended by 93/ 68/ EEC (CE
Marking). However, electrical noise or
intense electromagnetic fields in the vicinity
of the equipment may disturb the
measurement circuit. Measuring
instruments will also respond to unwanted
signals that may be present within the
measurement circuit. Users should exercise
care and take appropriate precautions to
avoid misleading results when making
measurements in the presence of electronic
interference.

Replacement Parts
TL36 

Test Lead Set

TL245 Test Lead Set
FP200  Fuse – 2 A / 1000 V (15XP and

35XP)

FP300 Fuse – 0.315A / 1000V (5XP)
H2-XR Magne-Grip

®

 Holster, clip, magnet,

and strap

XR-TA  Input Adapter for K-type

thermocouple

TP255 K type thermocouple

See www.metermantesttools.com for
detailed specifications of the Meterman
XP digital multimeters.

5XP Electrical Specifications

(at 23°C ± 5°C, < 75% R.H.)
DC VOLTS
Ranges: 200 mV, 2 V, 20 V, 200 V, 1000 V
Accuracy: ± (1.0% rdg + 1 dgt)

AC VOLTS (45 Hz to 500 Hz)
Ranges: 200 mV, 2V, 20 V, 200 V, 750 V
Accuracy: ± (1.5% rdg + 5 dgts)

DC CURRENT
Ranges: 200 µA, 2 mA, 20 mA, 200 mA,
± (1.5% rdg + 1 dgt)

AC CURRENT (45 Hz to 500 Hz)
Ranges: 200 µA, 2 mA, 20 mA, 200 mA,
± (2.0% rdg + 5 dgts)

RESISTANCE
Ranges: 200 

, 2 k

, 20 k

, 200 k

,

2 M

, 20 M

Accuracy: ± (1.0% rdg + 4 dgts) on 200 to
200 k

 ranges ± (1.5% rdg + 4 dgts) on

2 M

 range ± (3.0% rdg + 5 dgts) on

20 M

 range

CONTINUITY
Audible indication: 75 

±

 25 

DIODE TEST
Test current: 1.0 mA (approximate)
Accuracy: ± (1.5% rdg + 3 dgts)
Open circuit volts: 3.0 dc typical

Содержание 15XP

Страница 1: ...uga ON L4Z 1X9 Tel 905 890 7600 Fax 905 89 6866 European Correspondence Address Meterman Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspondence only no repair or replacement available from this address European customers please contact your distributor Visit www metermantesttools com for Catalog Application notes Product specifications Product manuals PN 2095408 October 20...

Страница 2: ...CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A 2A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA 2A A 1832 400 5 6 4 2 1 35XP 3 Document Dimensions 3 3125 X 5 125 ...

Страница 3: ...Continuity See Figure 6 5 Testing Diodes See Figure 7 5 Measuring Capacitance 35XP only See Figure 8 5 Measuring Temperature 35XP only See Figure 9 6 Measuring Frequency 35XP only See Figure 10 6 Measuring NCV Non Contact Voltage See Figure 11 6 Testing Battery Voltage 5XP only See Figure 12 6 Testing Logic Levels 15XP only See Figure 13 7 Additional Features 7 Input Test Lead Warning 7 MIN MAX Me...

Страница 4: ...ble equipment etc where only smaller transient overvoltages may occur and not for primary supply lines overhead lines and cable systems Do not exceed the maximum overload limits per function see specifications nor the limits marked on the instrument itself Never apply more than 1000 V dc 750 V ac rms between the test lead and earth ground Inspect the DMM test leads and accessories before every use...

Страница 5: ...e To return to autoranging press the RANGE button If RANGE still shows on the display autoranging is not available for the selected function Use autorange for all initial measurements Then when appropriate use the RANGE button to select and lock a range XWWarning To avoid electrical shock while manual ranging use the display annunciators to identify the actual range selected Correcting an Overload...

Страница 6: ... See Figure 3 1 Set the Function Switch to a current function µA mA or A 2 Select the desired RANGE 5XP only 3 Connect the test leads Red to mA or A Black to COM 4 Turn off power to the circuit being measured 5 Open the test circuit X to establish measurement points 6 Connect the test probes in series with the load to the measurement points 7 Turn on power to the circuit being measured 8 Read the ...

Страница 7: ...arge any capacitors that may influence the reading 5 Connect the test probes across the resistance 6 Listen for the tone that indicates continuity 40 Ω Testing Diodes See Figure 7 1 Set the Function Switch to G 2 Connect the test leads Red to E Black to COM 3 Turn off power to the circuit being measured 4 Free at least one end of the diode from the circuit 5 Connect the test probes across the diod...

Страница 8: ...ct Voltage See Figure 11 1 Range switch may be set to OFF or any function range 2 Test leads are not used for the NCV test 3 Press the NCV button The display goes blank a tone sounds and the green LED next to the NCV button on the front panel lights up to verify that the instrument is operational While pressing the button hold the top center of the meter z sensor location close to the conductor ci...

Страница 9: ...MAX button for less than 1 second will put the meter into a mode of displaying the maximum or minimum readings Each time the button is pressed the meter will cycle to the next display mode Press the MIN MAX button for more than 1 second to disable this feature Auto Power Off Models 15XP and 35XP only Auto Power Off is a battery saving feature that puts the meter into a sleep mode if the Function R...

Страница 10: ...15 A 1000V minimum interrupt rating 30 kA 6 3 x 32 mm Meterman FP300 WARRANTY The XP Series of Digital Multimeters are warranted against any defects of material or workmanship within a period of one 1 year following the date of purchase of the multimeter by the original purchaser or original user Any multimeter claimed to be defective during the warranty period should be returned with proof of pur...

Страница 11: ...r you Additionally in the United States and Canada In Warranty repair and replacement units can also be sent to a Meterman Test Tools Service Center see below for address Non Warranty Repairs and Replacement US and Canada Non warranty repairs in the United States and Canada should be sent to a Meterman Test Tools Service Center Call Meterman Test Tools or inquire at your point of purchase for curr...

Страница 12: ...tic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will also respond to unwanted signals that may be present within the measurement circuit Users should exercise care and take appropriate precautions to avoid misleading results when making measurements in the presence of electronic interference Replacement Parts TL36 Test Lead Set TL245 Test Lead ...

Страница 13: ...ranges 2 0 rdg 3 dgts on 2 A range AC CURRENT 45 Hz to 500 Hz Ranges 200 µA 2000 µA 20 mA 200 mA 2 A Accuracy 1 5 rdg 5 dgts on 200 µA to 200 mA ranges 2 5 rdg 5 dgts on 2 A range RESISTANCE Ranges 200 Ω 2 kΩ 20 kΩ 200 kΩ 2 MΩ 20 MΩ Accuracy 1 0 rdg 4 dgts on 200 to 2 MΩ ranges 3 0 rdg 5 dgts on 20 MΩ range CONTINUITY Audible indication Less than 25 Ω DIODE TEST Test current 1 2 mA approximate Acc...

Страница 14: ...acy 5 0 rdg 30 dgts on 4 nF ranges 5 0 rdg 5 dgts on 40 nF and 400 µF ranges 5 0 rdg 15 dgts on 4 mF range TEMPERATURE Ranges 20 C to 1000 C 4 F to 1832 F Accuracy 2 0 rdg 4 C 20 C to 10 C 1 0 rdg 3 C 10 C to 200 C 3 0 rdg 2 C 200 C to 1000 C 2 0 rdg 8 F 4 F to 50 F 1 0 rdg 6 F 50 F to 400 F 3 0 rdg 4 F 400 F to1832 F FREQUENCY Ranges 4 k 40 k 400 k 4 M 20 MHz Accuracy 0 1 rdg 3 dgts Sensitivity 1...

Страница 15: ...400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 4 4 1 2 V V 1 1 3 3 2 5 2 5 ...

Страница 16: ...00 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 4 4 7 5 6 4 7 5 6 3 3 1 3 1 A A 2 8 8 2 ...

Страница 17: ...00 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 4 6 5 5 3 4 6 1 5 6 2 1 2 7 3 ...

Страница 18: ...00 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A A mA 400 1000 A A 35XP RANGE NON CONTACT VOLTAGE HOLD V HZ V auto off C OFF F mA A A 1832 400 2 1 1 4 5 6 8 3 7 3 5 4 6 7 2 ...

Страница 19: ...17 10 1 3 2 2 4 K 1 9 3 4 5 ...

Страница 20: ...A V V OFF 200 m 2m 20m 1 5V 9V 200 BATT 5XP 2A MAX FUSED CAT 600V CAT 1000V CAT 300V COM A MAX 1000V 750V Hz Temp V Ω mA 2A A mA 400 1000 A A 35XP RANGE HOLD V HZ V auto off C OFF F mA A A 1832 400 11 1 2 3 2A MA X FUS ED CA T 600 V CA T 100 0V CA T 300 V CO M A MA X 100 0V 750 V Hz Tem p V Ω mA 2A 2A mA 400 100 0 A A 35XP RA NG E HO LD V HZ V auto off C OF F F mA 2A A 183 2 400 4 NCV 12 2 BATT 3 ...

Страница 21: ...19 ...

Страница 22: ...uga ON L4Z 1X9 Tel 905 890 7600 Fax 905 89 6866 European Correspondence Address Meterman Test Tools Europe P O Box 1186 5602 BD Eindhoven The Netherlands Correspondence only no repair or replacement available from this address European customers please contact your distributor Visit www metermantesttools com for Catalog Application notes Product specifications Product manuals PN 2095408 October 20...

Отзывы: