oxide  and  other  species  such  as  hydrogen  fluoride,  hydrogen  chloride,  hydrogen  sulphide,  carbon 
monoxide,  acetylene,  ammonia,  and  others.  The  absorption  line  can  be  scanned  under  software 
control and along with the electrical output of the gas cell the laser is an extremely useful tool in gas 
sensing and gas sensing system development.    

1.3

 

Visual and Operational Inspection 

The  Clarity  laser  is  packed  in  a  carton  designed  to  give  adequate  protection  during  shipment.  If  the 
outside of the shipping carton is damaged, notify your shipping department and carrier immediately. If 
the external packaging is not damaged carefully remove the contents. The contents should include: 

  Clarity laser 

  Power cord 

  Cal certificate and manual (manual may be emailed separately) 

  BNC rear interlock short 

1.4

 

Options and Accessories 

The  Clarity  laser  is  available  in  a  number  of  different  wavelengths;  check  with  factory  for  the  latest 
availability.  The  unique  architecture  used  allows  some  tuning  of  the  wavelength  while  still  being 
referenced to an unchanging molecular energy level. This allows for wavelengths to be offered that are 
exactly  on  an  ITU  DWDM  grid  location  (Note  only  available  on  some  grid  locations).The  wavelength 
option is designated by the center wavelength. For example: 

 

Clarity-PFR-1530 

Clarity-NLL-1542 

Clarity-NLL-1542-HP 

Explanation: 

PFR 

 

Stands for Precision Frequency Reference. 

NLL  

Stands for narrow linewidth laser. This includes PFR functionality 

1530 

 

The laser wavelength in nm. Standard versions are: 

1530nm

 locked to the P9 line of C12 acetylene 

1542nm 

locked to the P16 line of C13 acetylene 

1312nm

 locked to the P3 line of hydrogen fluoride 

[other wavelengths available 

– consult factory] 

HP   

 

High power option for the NLL, includes PM fiber output at >30mW 

 

1.5

 

Operating Environment 

The Clarity laser should be used in an area which satisfies the following conditions: 

  Ambient Temperature: 0

C to 50

C (operating temperature) 

  Relative Humidity: <85% non-condensing 

  Low  noise  area.  While  the  Clarity  laser  is  designed  for  noise  immunity  the  best 

performance will be achieved in a low EMF area. Please use a noise line filter in an area 
where high noise is unavoidable 

Содержание Clarity Series

Страница 1: ...Molecular Absorption Line Operating Manual 0700 M Rev C Copyright Wavelength References 2012 All Rights Reserved Wavelength References 1800 SW 3rd Street 140 Corvallis OR 97333 Voice 541 738 0528 Fax...

Страница 2: ...duct to Customer if the shipment is within the United States or Canada Customer shall be responsible for paying all shipping charges duties taxes and any other charges for products returned to other l...

Страница 3: ...ed 7 4 1 1 Password 7 4 1 2 Initial Calibration 7 4 1 3 Interlock 7 4 2 Front Panel Operation 7 4 2 1 Menu 7 4 2 2 LOCK MODE 9 4 2 3 CAL 9 4 2 4 ACTIVE 9 4 3 Front panel display 9 4 3 1 Frequency and...

Страница 4: ...lication for the Clarity laser is to provide an extremely accurate and stable wavelength standard in the DWDM band This can be used as a calibration source for an OSA optical spectrum analyzer or to v...

Страница 5: ...wavelength while still being referenced to an unchanging molecular energy level This allows for wavelengths to be offered that are exactly on an ITU DWDM grid location Note only available on some gri...

Страница 6: ...difficulties Another possibility investigated was frequency doubling 1500 1560nm light to probe atomic transition in the 750 780nm area This approach requires complicated and expensive apparatus Mole...

Страница 7: ...ant temperature There is essentially no long term drift due to the gas line reference The tube construction uses a low melting temperature glass that results in a true hermetic seal The line center lo...

Страница 8: ...Take care to clean all fiber connections to the laser well before making connection This will help keep the internal ferrule clean and undamaged If the internal ferrule is dirty you can attempt to cl...

Страница 9: ...s the command 4 1 2 Initial Calibration Our high power 30mW lasers include a warm up cycle when first turned on After the laser is first turned and the ACTIVE button is pressed the front panel display...

Страница 10: ...n press NEXT The front panel will display Laser Password and the laser will be off To re enable the laser press and hold either directional button to scroll to the correct password and then press NEXT...

Страница 11: ...weeps the laser across a limited spectral range approx 150 GHz and identifies the position of the gas line s peak absorption This information is used in the instrument s locking routines Should you pr...

Страница 12: ...t was opened then closed 2 the instrument s locking routine exceeded its dynamic range and had to re center to the absorption line This can occur if the instrument s temperature changes suddenly To cl...

Страница 13: ...e used as a diagnostic or in gas sensing system experiments where the laser is scanning the line to provide an indication of the exact timing and scanning of the absorption feature 3 RS 232 The Clarit...

Страница 14: ...e Boolean 0 1 off on Sets laser enabled disabled state SOURce STATe Queries laser enabled disabled state SOURce WAVElength xxxx xxxxxnm Queries target wavelength System Subsystem SYSTem ERRor NEXT Que...

Страница 15: ...EEE 488 2 Requirements CLS Clear status command ESE numeric value Standard event status register ESE IDN Identification query OPC Operation complete enable OPC RST Reset command STB Read status byte q...

Страница 16: ...cted 220 Parameter error No entry in list to retrieve number list or channel list 221 Settings conflict Check rear interlock Settings Conflict Check rear interlock 222 Data out of range Data out of ra...

Страница 17: ...equency standard locked to a physical constant We rely on published wavelength data from HITRAN or NIST where available Wavelength Reproducibility 2 MHz 600Khz RMS Typical At 25 C 5 C after self calib...

Страница 18: ...classification may be shipped by any customary means 8 Bibliography 1 F Nez F Biraben R Felder and Y Millerioux Opt Comm 102 432 438 1993 2 K Nakagawa M de Labachelerie Y Awaji and M Kourogi Accurate...

Отзывы: