5.  Setup Monitors

Attach a DVM or other voltage monitoring device to the Current Monitor (Pin 10) and Limit Monitor 
(Pin 11).  Reference to Monitor Ground (Pin 12).  Transfer functions are:

Examples:
QCL500 - 2.5 V on the Current Monitor pin, equivalent current is (2.5 V * 0.2 A / V) or 500 mA.  
QCL1000 - 1.8 V on the Current Monitor pin, equivalent current is (1.8 V * 0.4 A / V) or 0.72 A.
QCL1500 - 2.0 V on the Current Monitor pin, equivalent current is (2.0 V * 0.6 A / V) or 1.2 A.

QCL500 - 5 V on the Limit Monitor pin, equivalent limit current is (5 V * 0.09 A / V) or 450 mA.
QCL1000 - 2.5 V on the Limit Monitor pin, equivalent limit current is (2.5 V * 0.19 A / V) or 470 mA.
QCL1500 - 3.2 V on the Limit Monitor pin, equivalent limit current is (3.25 V * 0.28 A / V) or 910 mA.

6.  Power up the unit

Turn on power to the DC supplies.  Flip the POWER ON switch to ON.  The POWER ON LED lights.  If 
the LED does not power on, review REMOTE vs. LOCAL switch settings (Step 4) and the state of the 
REMOTE POWER ON signal (Pin 4 referenced to Pin 3).  

7.  Set Current Limit & Zero Output Current

Choose Current Limit current to be above the normal setpoint, but below the damage threshold of 
the QCL.  Monitor the Current Limit (between Pin 11 & 12).  Rotate the Current Limit Setpoint trimpot 
clockwise to increase the limit current setting.  Example conversions:  

QCL500 - Want 425 mA Limit, equivalent voltage is (0.425 A / 0.09 A / V) or 4.7 V.
QCL1000 - Want 750 mA Limit, equivalent voltage is (0.750 A / 0.19 A / V) or 3.9 V.
QCL1500 - Want 1.5 A Limit, equivalent voltage is (1.5 A / * 0.28 A / V) or 5.3 V.

Rotate the SET trimpot on the front panel completely counter clockwise to zero the output current 
setpoint.

OPERATING PROCEDURES, continued

Model

QCL500

QCL1000

QCL1500

Limit MON

0.09 A / V

0.19 A / V

0.28 A / V

Current MON

0.2 A / V

0.4 A / V

0.6 A / V

•  Power On:  Switch and indicator
•  Enable Current: Switch and indicator
•  Over Temperature Fault indicator
•  Current Setpoint trimpot 
•  Current Limit Setpoint trimpot

QCL1500

01A105409201

Covered by U.S. 
Patents 6,696,887; 
6,867,644 and 
7,176,755

Содержание QCL1000 LAB

Страница 1: ...587 4910 or quietqcl teamwavelength com 1 Covered by U S Patents 6 696 887 6 867 644 and 7 176 755 Licensed from Battelle Memorial Institute Minimize spectral drift center wavelength jitter and linew...

Страница 2: ...rt number TCH35P10R0JE ND 10 50 W metal film resistor QCL Pin 19 QCL Pin 20 Internally grounded Figure 2 Example Test Load NOTE At 2 A output current will be compliance voltage limited with a 10 resis...

Страница 3: ...nt Inrush current requirement 6 TYP MIN MAX UNITS TEST CONDITIONS PARAMETER TAMBIENT 25 C TAMBIENT 25 C 24 V in to full scale to full scale large signal 90 full scale to full scale at 500 kHz IOUT 100...

Страница 4: ...er Function Remote Enable Disable Remote Power On Status Outputs MONITOR Current Monitor Range Current Monitor Bandwidth 3 dB Current Monitor stability Accuracy Current Monitor to Actual Limit Monitor...

Страница 5: ...ed Model QCL500 QCL1000 QCL1500 RMS noise A 100 kHz BW 0 4 0 7 1 0 Average Current Noise Density nA SHz 1 0 2 0 3 0 How noise measurements are made is detailed in WEI s Low Noise Measurement Protocol...

Страница 6: ...FICATIONS continued There is a slight 2 overshoot when the driver is operating in current limit with the limit value set well below the setpoint We recommend that you operate with the limit above the...

Страница 7: ...or Analog Input Voltage Not designed for high current return Ground can be used to ground chassis Positive DC power to the unit typically 24 V Negative DC power to the unit typically 24 V GROUND desig...

Страница 8: ...f this limit Online at http www teamwavelength com support calculator soa soald php choose the model of QCL driver you will be using Enter your power supply and QCL characteristics to determine intern...

Страница 9: ...sing the REMOTE inputs wire them now TTL compatible up to 12 V Pin 3 Remote Power On Pin 4 Remote Power On Pin 5 Remote Enable Pin 6 Remote Enable STATE Local Only All switches to right or ON FACTORY...

Страница 10: ...g a power supply with the chassis not tied to ground WEI offers switching power supplies PWRPAK 24V a 20 pin cable WCB311 and a power supply kit WCB312 to simplify setup Twist wire pairs to minimize n...

Страница 11: ...hts If the LED does not power on review REMOTE vs LOCAL switch settings Step 4 and the state of the REMOTE POWER ON signal Pin 4 referenced to Pin 3 7 Set Current Limit Zero Output Current Choose Curr...

Страница 12: ...off ramp is 5 5 sec Flip the POWER switch to the OFF position The POWER LED goes out Turn off the power to the dual DC power supplies 10 Remove test load and wire in a Quantum Cascade Laser Wire the...

Страница 13: ...or 50 Hz peaks Pin 15 on the 20 pin connector is ground It can be wired to a screw on the chassis to the mounting slots etc Note that despite the peak at 60 Hz 0 ot 100 kHz RMS noise for the two graph...

Страница 14: ...PIN Ground Pin 12 10 k The STATUS pins output 12 V up to 25 mA With this typical hookup we assume 1 5 V forward voltage across LED and 1 mA drive current 12 V 1 5 V 1 mA 10 k Minimize cable length to...

Страница 15: ...20 CABLE 4 cond 4 cond 4 cond 4 cond 6 cond 6 cond 6 cond 6 cond 6 cond 6 cond 2 cond 2 cond 1 cond 3 cond 3 cond 3 cond 2 cond 2 cond BLK WHT GRN RED BLU GRN ORG WHT RED BLK RED BLK BLK RED WHT BLK...

Страница 16: ...00 Low Noise Quantum Cascade Laser Drivers PAGE 16 QCL1500 00400 D MECHANICAL SPECIFICATIONS 2 32 6 00 28 6 55 4 00 50 5 00 5 50 75 R 141 281 233 TYP 59 062 Dimensions are in inches mm Tolerance is 5...

Страница 17: ...ver needs to be mounted by the baseplate mounting holes not supported by the solder joints The solder joints are not meant for mechanical support To use 8 screws instead of 1 4 20s for mounting a shou...

Страница 18: ...NOTE C NOTES NAME PLATE 7 3 10 1 VOLTAGE ADJUSTMENT 5 M3 5 SEE NOTE B LED 82 1 8 15 MORE THAN 6m m SCREWS MUST NOT PROTRUDE INTO POWER SUPPLY BY C M3 TAPPED HOLES 2 FOR CUSTOMER CHASSIS MOUNTING AC L...

Страница 19: ...e example SOA determination Device QCL1500 V 24 Volts VQCL 10 Volts IQCL 1 25 Amp Follow these steps 1 Determine the maximum voltage drop across the driver V VQCL and mark on the X axis Example 24 V 1...