Страница 4: ...e analog interface to disable an LFI 4500 s laser diode driver output if the temperature controller output is turned off for any reason Interlock any combination of LFI 3500 series temperature control...

Страница 5: ...D sub rec s Both ends connectorized All cables are 1 meter long Rack Mount Kits Selection Guide Model Description RMK 502 Rack mount kit for two LFI modules Includes two 5 1 4 panels RMK 503 Rack mou...

Страница 6: ...1 10 8 CONFIGURATION Sensor setting Thermistor 100 A reference current Loop Direction NTC for thermistors PID Constants P 32 I 1 second D OFF Temperature Limit HIGH Front Panel LIMITS Temperature Limi...

Страница 7: ..._____________ 20 Operation with RTDs________________________ 22 Set the Thermoelectric Current Limit __________ 25 Set the Temperature Limit ___________________ 26 Further Information_________________...

Страница 8: ...t could possibly cause bodily injury or death ATTENTION Ce symbole signale une proc dure ou des conditions dangereuses pouvant entra ner des blessures corporelles ou la mort VORSICHT Nichtbeachtung de...

Страница 9: ...are removed NO USER SERVICEABLE PARTS INSIDE ATTENTION Risque d Electrocution Ne pas mettre sous tension cet instrument si le bo tier est endommag ou si les capots ou les faces sont retir s Aucun com...

Страница 10: ......

Страница 11: ...nt is limited to the adjustment set by the Limit Current Trimpot This condition will remain until the current required to achieve temperature falls below the Limit Current Trimpot setting This conditi...

Страница 12: ......

Страница 13: ...witch be in the ON position Switch 6 Integral Term This switch configures the integral time constant to be either one or ten seconds In the OFF position the integral time constant is one second In the...

Страница 14: ...ns 5 6 Err Err Connect to LFI 4500 laser diode drivers Pin 7 Error Detect Error status Pin 8 On Off Detect On Off status Pin 9 Common Common for pins 7 8 10 Pin 10 Remote On Off Remote On Off control...

Страница 15: ...correct power line fuse is installed and the power line voltage setting is correct The factory sets the following defaults based on the country of destination for initial shipment Destination Voltage...

Страница 16: ......

Страница 17: ......

Страница 18: ......

Страница 19: ...perature Setpoint Rotate the Display Select Switch to the LINEAR position The value displayed will be the ambient temperature in Kelvin LINEAR IC Sensor Display Setting Press and hold the Display Set...

Страница 20: ......

Страница 21: ...e ambient temperature in Kelvin LINEAR IC Sensor Display Setting Press and hold the Display Set Button Rotate the Adjust Potentiometer until the desired setpoint temperature is displayed Step 8 Enable...

Страница 22: ......

Страница 23: ...ature NOTE Ignore the decimal point on the 3 digit display and the resistance of the RTD is indicated in ohms RTD Display Setting Press and hold the Display Set Button Rotate the Adjust Potentiometer...

Страница 24: ...24...

Страница 25: ......

Страница 26: ......

Страница 27: ......

Страница 28: ...e range than the thermistor but the thermistor will be more sensitive to temperature changes PID PROCESSOR The PID processor section consists of a proportional gain amplifier an integrator and a diffe...

Страница 29: ......

Страница 30: ...above ambient OFF ON 2 3 4 5 6 7 9 1 10 8 P 64 I 10 sec D 250 msec Typical Settings for AD590 LM335 Small Thermal Load OFF ON 2 3 4 5 6 7 9 1 10 8 P 32 I 1 sec D none Medium Thermal Load OFF ON 2 3 4...

Страница 31: ...d to the output connector This condition does not shut down the output current and is only used to indicate a possible output connection problem Sensor Error LED This LED lights red whenever the main...

Страница 32: ...er amplifier to boost the power output of the controller In this case the thermoelectric is connected to the booster power amplifier while the LFI controller provides the PID temperature control This...

Страница 33: ...ransfer function is 1 A V This output is internally limited to 10 mA maximum output current Pins 13 15 Temperature Setpoint Monitor Common This pin provides a buffered measurement of the temperature s...

Страница 34: ...50 or 60 Hz switch selectable on rear panel Maximum AC Input 250VAC Power Consumption 160VA Peak Size W x D x H 106 mm x 300 mm x 163 mm 4 25 x 12 x 6 5 Weight 4 5 kg 10 lb Operating Temperature 0 to...

Страница 35: ...k 0 19 99 k 0 19 99 k 10 A range 0 199 9 k 0 199 9 k 0 199 9 k 0 199 9 k Linear IC Sensors 0 1999 Kelvin 0 1999 Kelvin 0 1999 Kelvin 0 1999 Kelvin TE Current 2 50 A 2 50 A 5 0 A 5 0 A TE Voltage 9 99...

Страница 36: ...36 MECHANICAL SPECIFICATIONS...