© 2018
20
LDTC0520 / LDTC1020 LASER DIODE AND TEMPERATURE CONTROLLER
AN ALTERNATE METHOD FOR SETTING THE
LASER DRIVER CURRENT LIMIT
The laser drive current can be set while using the test load,
as recommended on
, or it can be set while
monitoring an internal test point. Using this method, it is
not necessary to connect a test load or switch on the driver
output, but the cover must be removed from the LDTC.
Refer to your laser diode datasheet to find the maximum
recommended forward current, I
LIM
. Then calculate the test
point voltage value for the current limit setting. The transfer
, and varies according to the
controller model number.
V
LD I LIM
=
I
LIM
(9)
Transfer Function
Remove the two screws securing the cover of the LDTCxx20,
lift off the cover, and locate the test point above the I LIM
for
Rev A & B only. Connect the positive lead of the DVM to the
test point, and the negative lead to one of the ground pins on
the J2 connector (4, 6, 12).
Connect the laser driver power supply and switch it on.
ILIM
J2
1
12
Figure 28. Current Limit Test Point and Trimpot —
Rev C only
Adjust the I LIM trimpot until the voltage displayed on the
multimeter matches the V
LD I LIM
value calculated above.
ILIM
J2
1
12
Figure 29. Current Limit Test Point and Trimpot —
Rev A & B only