CARDIO WAVE EXCITE Class & Trend: Service & maintenance manual- rev.2.0
Pagina 6.35
6.14.
THERE IS NO HEART RATE SIGNAL
6.14.1.
SALUTRON
TELEMETRIC RECEIVER
The machine displays this error if:
1.
electromagnetic noise saturate the HR receiver which does not display any value due to a
specific SW filter which cut every signal, greater than 220 bpm;
2.
the receiver is not working properly;
3.
the receiver is not supplied by the display board.
THERE IS NO HR
SIGNAL
Is the supply v oltage correct at
the input to the receiv er?
Is the supply v oltage correct at
the output of the display board?
Replace the receiv er
Replace CU223 cable
2
3
YES
NO
YES
Replace display board
NO
Carry out the
troubleshooting procedure
"THE TELEMETRIC
SIGNAL IS INCORRECT"
1
Содержание CLASS & TREND
Страница 1: ...SERVICE MAINTENANCE MANUAL REV 2 0 ...
Страница 2: ......
Страница 4: ......
Страница 10: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 1 6 Page intentionally left blank ...
Страница 32: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 2 20 Page intentionally left blank ...
Страница 44: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 3 12 Page intentionally left blank ...
Страница 50: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 4 6 LIBYA DVB T ...
Страница 51: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 4 7 Page intentionally left blank ...
Страница 52: ......
Страница 58: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 5 6 Page intentionally left blank ...
Страница 100: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 6 42 Page intentionally left blank ...
Страница 144: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 7 44 Page intentionally left blank ...
Страница 202: ...CARDIO WAVE EXCITE Class Trend Service maintenance manual rev 2 0 Pagina 9 50 Page intentionally left blank ...
Страница 213: ......