5.9 Firmware Update
Software for the CSM can be updated using an SD card. While the
software is being updated, it is important to not turn off power to
the instrument. Once the software update has been complete, the
instrument will require a reboot. To ensure the software update was
successful, you can view the software version in the About
Device menu.
To update the software on the CSM:
1. Press
Firmware Update
on the Configuration Menu screen.
2. Insert an SD card containing the software update into the SD slot on
the display board located on the electrical panel door.
Press
Accept
.
3. Press which software option you wish to update:
Update Main
Display
,
Update Main UI
,
Update Main Measurement
.
4. After pressing the update option, the firmware will begin to update.
A status update will appear on the screen to display the progress of
the update. Once the update is complete, a restart of the instrument
is required.
Do not remove the SD card or turn power off while the software
update is in progress. Doing so can cause an
unrecoverable failure.
NOTICE
5.10 Copying FRAM data to SD Card
The following information from the CSM can be transferred to an SD
card in CSV file format.
• Readings
• Error and warning logs with date and time stamps
• All user settings
• Calibration data
To transfer data from the CSM to an SD card:
1. Press
Copy FRAM Data to SD Card
on the Configuration Menu
screen.
2. Insert an SD card into the SD slot on the display board located on
the door of the electrical panel and press
NEXT
.
3. After pressing NEXT, the data will being copying to the SD card. A
status update will appear on the screen stating Copying Records to
SD Card in Progress... Once the data has been copied to the SD
card, a completion update will appear on the screen.
Display Board Located on the Electrical Panel Door
SD Card Slot on the Display Board
Display Board Located on the Electrical Panel Door
SD Card Slot on the Display Board