7.0 Instrument Configuration (CONFIG mode)
The instrument has been designed to provide the ability to customize the instrument
according to needs at any time during normal operation. This mode has been split into
sub-menus to facilitate instrument configuration. This section describes how to use each
of the sub-menus to configure the instrument. While in the configuration mode, the
instrument has a time-out feature that automatically returns the system operation to the
AUTO
mode after a fifteen (15) minute period of no button pushes.
Enter the
CONFIG
mode of the instrument by pressing the
MODE/EXIT
button until the
arrow beside
CONFIG
is illuminated, then press the
↵
button to scroll through the sub-
menus.
To exit the CONFIG mode, press the
MODE/EXIT
button any changes that were
made will be saved.
7.1 Setting the 4-20 mA Output
The first configuration selection is
4-20
for the 4-20 mA
output. Select the either
On
or
OFF
using the
and
buttons. Once the desired output has been set, press the
button to accept it. The next prompts will depend on
the output selected. Also see sections 7.13 & 7.14.
If the 4-20 mA output was turned
On
, prompts to set the
4mA
(4MA)
and 20mA
(20MA)
chlorine limits will be
displayed. There will also be a menu to adjust the error
level (ERLV). The first prompt will be the chlorine limit
assigned to the 4 mA output level:
Select the chlorine level to assign to the
4MA
using the
and
buttons.
Once the desired level has been set, press the
button to accept it.
The 4MA can be set higher than 20MA level to invert the output current if required.
This may be required to control a dosing pump.
The next prompt will be the chlorine level assigned to the
20MA. Select the chlorine level using the
and
buttons. Once the desired level has been set, press the
button to accept it.
7.2 Configuring the Error Level
In case of an error in the CLX-XT, the 4-20 mA reading
can be used to indicate a problem by sending the current
to either 4.00 mA, 2.00 mA or 0 mA. The factory
default setting is OFF. Select the desired ERLV by using
the
and
buttons then press the
button to accept
the desired error response.