Watts HF scientific CLX-Ex2 Series Скачать руководство пользователя страница 13

13

Please note that these operating screens cannot be seen 

except through the door window with the cabinet door 

closed or in maintenance bypass mode. See section 5.3 

for more information on this.

If changes to the configuration have to be performed, you will need to 

enter the maintenance bypass mode see section 5.3 for more information.
The instrument has been designed to provide the ability to customize the 

instrument according to needs at any time during normal operation. This 

mode has been split into sub-menus to facilitate instrument configuration. 

This section describes how to use each of the sub-menus to configure the 

instrument. While in the configuration mode, the instrument has a time-out 

feature that automatically returns the system operation to the 

AUTO

 mode 

after a fifteen (15) minute period of no button pushes.
Enter the 

CONFIG

 mode of the instrument by pressing the 

MODE/EXIT

 

button until the arrow beside 

CONFIG

 is illuminated, then press the 

  

button to scroll through the sub- menus.

8.1 Setting the 4-20 mA Output

The first configuration selection is 

4-20

 for the 4-20 mA 

output. Select the either 

On

 or 

OFF

 using the 

 and 

 

buttons. Once the desired output has been set, press the 

  button to accept it. The next prompts will depend on the 

output selected. Also see sections 8.12 and 8.13.

If the 4-20 mA output was turned 

On

, prompts to set the 4mA 

(4MA)

 and 

20mA 

(20MA)

 chlorine limits will be displayed. There will also be a menu 

to adjust the error level (ERLV). The first prompt will be the chlorine limit 

assigned to the 4 mA output level:
Select the chlorine level to assign to the 

4MA

 using the 

 and 

 buttons.

Once the desired level has been set, press the 

  button to accept it.

NOTICE

To exit the CONFIG mode, press the MODE/EXIT button any 

changes that were made will be saved.

NOTICE

The 4MA can be set higher than 20MA level to invert the output 

current if required. This may be required to control a dosing pump.

The next prompt will be the chlorine level assigned to the 20MA. Select 

the chlorine level using the 

 and 

 buttons. Once the desired level 

has been set, press the 

  button to accept it.

8.2 Configuring the Error Level

In case of an error in the CLX-Ex2, the 4-20 mA reading can be used to 

indicate a problem by sending the current to either 4.00 mA, 2.00 mA or 0 

mA. The factory default setting is OFF. Select the desired ERLV by using the 

 

and 

 buttons then press the 

  button to accept the desired error response.

8.3 Configuring the RS–485 Port

The instrument is equipped with an RS 485 port which operates in 

Simple bus or Modbus. Prompts will appear for setting the baud rate, 

the address and the Modbus transmission mode (RTU or ASCII).

Select the correct baud rate (1200, 2400, 4800, 9600, or 19200) for 

operation of the I/O port by pressing the 

 or 

 buttons to change the 

displayed baud rate.
Press the 

  button to continue on and select the desired instrument 

address using the 

 or 

 buttons. Once the selection is satisfactory, 

press the 

  button.

To use the Modbus mode, select 

ASCII

 or 

RTU

. See 

section 9.1.2.

Only access when the atmosphere has been tested and known to 

be safe. 

WARNING

!

8.0  Instrument Configuration 

(CONFIG mode)

Содержание HF scientific CLX-Ex2 Series

Страница 1: ...ded Reagent Life Model CLX Ex2 28034 IOM HF CLX Ex2 28920 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious per...

Страница 2: ...7 1 Slope gain Calibration Procedure 12 7 2 Zero offset Calibration Procedure 12 7 3 Restore Factory Settings 12 8 0 Instrument Configuration CONFIG mode 13 8 1 Setting the 4 20 mA 13 8 2 Configuring...

Страница 3: ...stances icons or symbols have been added to give further clarification to the instructions Refer to the Table of Contents to easily find specific topics and to learn about unfamiliar terms This manual...

Страница 4: ...d Communications Port Bi directional RS 485 with Modbus Water Pressure Integral pressure regulator 0 34 bar 5 0 PSI to 10 3 bar 150 PSI Flow Rate to Waste 200 400 ml min Operating Temperature 0 C 55 C...

Страница 5: ...nters the option or mode that is highlighted or chosen The and buttons are used to change settings The PRIME and SERVICE buttons are dedicated controls The PRIME will start 75 reagent pump pulses to p...

Страница 6: ...LX Ex2 should be removed If the CLX Ex2 is to be turned off it is recommended that the instrument be placed in SERVICE mode before removing power This ensures that the cuvette is emptied and the flow...

Страница 7: ...more information on this feature refer to sections 4 4 4 Remote Standby Connections and 6 2 Routine Measurement with Remote Standby 3 5 The Reagent Cooler A reagent cooler is provided for the 125 ml i...

Страница 8: ...l terminals are designed to accept wires in the range of 14 28 AWG All wires should be stripped to a length of 6 mm For line power connections refer to the companion manual Catalog No 28749 4 4 1 RS 4...

Страница 9: ...on of the main enclosure with clean dry air Should this air supply pressure fail the PRESSURE lamp on the Air Sentinel unit will turn RED and power to the rest of the instrument is disconnected 5 1 St...

Страница 10: ...chlorine of process water on line The chlorine value of the process water is usually reported in milligrams per Liter mg L these units are equivalent to Parts Per Million PPM The CLX Ex2 has two avail...

Страница 11: ...is the number to be changed Use the or arrows to select the first of the three numbers in the code and then press the button to accept the first number of the code Now enter the second number in the...

Страница 12: ...size of the change that can be made to a current reading The upper limit is the current reading times 1 5 The lower limit is the current reading divided by 1 5 Ensure a reading is posted to the displ...

Страница 13: ...vel ERLV The first prompt will be the chlorine limit assigned to the 4 mA output level Select the chlorine level to assign to the 4MA using the and buttons Once the desired level has been set press th...

Страница 14: ...he desired UNIT using the and buttons and press the button to accept it 8 8 Averaging and Filtering The CLX Ex2 can display and output averaged readings to help smooth out the response and eliminate l...

Страница 15: ...generally not used in ballast water applications and should be set to OFF The instrument requires 110 seconds to complete its normal operations when the WCON is turned On at the factory cycle time the...

Страница 16: ...Additional Features and Options 9 1 RS 485 Outputs The CLX Ex2 has the capability to operate in two different RS 485 modes Included are a simple communication mode and Modbus communications Both mode...

Страница 17: ...tor can correct and the unit must be returned to the factory for service see section 10 6 These failures consist of failures in the CPU A D EEPROM or other devices internal to the instrument The queue...

Страница 18: ...pare Buffer Cap Assembly 1 Spare Indicator Cap Assembly 8 Pump Tubes 1 Spare Cuvette Two replacement sets of reagent cap tubing assemblies are supplied with the CLX Ex2 one for the buffer and one for...

Страница 19: ...rning the knurled top clockwise until the cuvette is held securely 5 Check the drawing on the following page to ensure correct installation 6 Return to operation as described T Strainer Cleaning The T...

Страница 20: ...s 11 4 Check Valve Flushing Kit Sometimes upon initial commissioning the check valves stick and require manual priming This should not be needed after commissioning Be careful when using this kit to u...

Страница 21: ...replacement parts Accessory Catalog Number Photo Reagent Kit Total Chlorine 12 month supply 09991 Air Sentinel Purge Control Operating Manual 28749 N A Operating Manual CLX Ex2 28920 N A Tubing Cuvet...

Страница 22: ...es and replacement parts To order any accessory or replacement part please contact the HF scientific Customer Service Department If for any reason technical assistance is needed regarding this instrum...

Страница 23: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 24: ...dy for breach of warranty and the Company shall not be responsible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing o...

Отзывы: