background image

4

3.4 Removing the Cyclops Vent

Tools and Supplies Required:

14mm (or 

9

16

") open end wrench

23mm (or 

29

32

") closed end wrench

Steps to remove the Vent:
1.  Remove the connection to the external interconnection stainless 

steel tube if this has not already been done. This requires a 14mm 

(or 

9

16

") open end wrench.

2.  From inside the enclosure use a 23mm (

29

32

") closed end wrench to 

remove the nut and washer, then remove the vent.

3.5 The Information Sticker

When replacing the Cyclops with the Air Sentinel one of the four front 

door yellow stickers will have incorrect information. The kit includes a 

replacement sticker to be used. The stickers are exceedingly difficult to 

remove. To make this easier, we recommend that this sticker be placed 

over the existing sticker that showed the Minimum and Maximum 

Overpressure values.

3.2 Temperature Controller Finger Guard

To gain better access the back nuts the Temperature Controller Finger 

Guard should be removed.

Tools Required:

A small size Philips screwdriver
Located above the Electronic PCB and chassis is a large finger guard 

and Temperature Controller cover. This is easily removed by loosening 

the four screws. Use care not to lose these screws as they are quite 

small.
After replacement of the controller, be sure the reattached the cover 

and finger guard.

Remove all power from the system before starting removal to 

avoid risk of electrocution.

WARNING

!

3.3 Cyclops Removal

Tools Required:

14mm (or 

9

16

") open end wrench

11mm (or 

7

16

") combination wrench and socket wrenches

Small flathead screwdriver 2.5 - 4mm (or 

1

8

 - 

5

32

")

Steps:

1.  Remove the power from the system.
2.  Remove the wire connections at terminals 4, 5 and 6 of the Cyclops.
3.  Remove the external interconnection stainless steel tube. Undo  

the connection at both ends. This requires a 14mm (or 

9

16

") open 

end wrench.

4.  Remove the Chassis. See section 1.1 above.
5.  Remove the 10 perimeter nuts flat washers and lock washers. This 

will require 11mm (or 

7

16

") tools. Often a combination wrench and a 

socket set work the best.

6.  Once all the hardware is removed you may need to give small tug as 

the rubber gasket can stick to the enclosure.

7.  After the hardware is removed, there is a small safety cover and a 

DIN rail containing a terminal. The safety cover can be discarded, 

but the Din rail will need to be replaced when installing the  

Air Sentinel.

Use care as the Cyclops weighs over 2.3 kg (5 lbs).

CAUTION

!

Содержание HF scientific Air Sentinel H-28136

Страница 1: ...lacement Kit 29207 WARNING Read this Manual BEFORE using this equipment Failure to read and follow all safety and use information can result in death serious personal injury property damage or damage to the equipment Keep this Manual for future reference IOM HF Air Sentinel Replacement Kit 29207 Rev 4 ...

Страница 2: ...ir supplies To prevent hand injury is recommended that safety gloves be worn 2 2 Safety 2 3 Hardware As there will be quite a bit of hardware removed it may be handy to have a small container to prevent losing any parts during the service work This is a safety alert symbol The safety alert symbol is shown alone or used with a signal word DANGER WARNING or CAUTION a pictorial and or a safety messag...

Страница 3: ... and CLX Ex2 3 1 The Chassis Removal and Replacement To gain access to the Cyclops mounting hardware it is first required to remove the chassis Tools Required A medium size Philips screwdriver The Electronic PCB and Chassis assembly is the complete system as shown to the right This includes the LCD keypad power supplies and cables Steps 1 Remove electrical connectors to Optics PCB 2 places the int...

Страница 4: ...r cover This is easily removed by loosening the four screws Use care not to lose these screws as they are quite small After replacement of the controller be sure the reattached the cover and finger guard Remove all power from the system before starting removal to avoid risk of electrocution WARNING 3 3 Cyclops Removal Tools Required 14mm or 9 16 open end wrench 11mm or 7 16 combination wrench and ...

Страница 5: ...all are tight 4 3 Air Vent Connection Tools Required Knife or Scissors The vent from the top of the optics assembly previously had no final connection A check valve was used to prevent water from entering the enclosure This vent must now connect to a fitting on the Air Sentinel through the same check valve The tube is shortened by about 50 60mm This tube will be placed on the output of the check v...

Страница 6: ...own in section 2 1 2 Remove the Temperature Controller Finger Guard as described in section 2 2 3 Remove the existing air vent connection See section 3 3 4 Remove the Mains and Power Out connections terminal blocks using the flat blade screwdriver It is not necessary to wires from the terminal blocks unless these terminal blocks have been damaged See section 3 4 5 Remove the 10 Washer Locknuts 6 R...

Страница 7: ... _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 8: ...ible for any incidental special or consequential damages including without limitation lost profits or the cost of repairing or replacing other property which is damaged if this product does not work properly other costs resulting from labor charges delays vandalism negligence fouling caused by foreign material damage from adverse water conditions chemical or any other circumstances over which the ...

Отзывы: