3.0 Theory of Operation
3.1 The Measurement
The CLX-Ex2 has two solenoid valves, one for sample water (FLOW)
and one for draining of the cuvette (PURGE). A third solenoid, along
with four check valves forms a reagent pump. Sample water flow is
controlled by the FLOW solenoid valve. The PURGE solenoid valve is
used to empty the cuvette in the measurement chamber.
The measurement chamber consists of a sample inlet, a purge drain,
and an overflow. The reagent is added from the check valves integrated
into the lower portion. A green LED provides the 515 nm source
lamp, a red LED is used for sample level and flow measurement. A
single detector is located 180º from the green LED. A replaceable
glass cuvette separates the LEDs from the detector and maintains
the measurement path length. Sample water flows in the inlet at the
bottom, through the measurement cuvette and out through an overflow
drain. This flow is used to both fill the cuvette and flush the system.
The reagents are dispensed from two replaceable bottles. One bottle
has a buffer to control the pH; the second has an indicator that contains
the DPD, which produces color when chlorine is present in the sample.
The degree of color is dependent on the amount of chlorine in the
sample water.
The measurement chamber is open to view operations. A white LED
backlights the chamber for a clearer view. The white LED will flash to
attract attention in the case of a warning or failure. Most warnings and
failures are also displayed on the screen. To prevent interference, the
white LED is turned off during measurements.
During normal operation the CLX-Ex2 will run through a timed cycle.
A simplified cycle will consist of the following sequences:
• Flushing – continuous sample flow
• Purging – PURGE valve opens
• Zeroing – no flow with cuvette full
• Adding Reagents – one pulse of the reagent pump
• Mixing with sample – sample flow pulses in
• Reading resulting sample – no flow with cuvette full
• Purging – PURGE valve opens to remove reacted sample
The cycle above is simplified and does not describe all the actions
and testing that occurs. The CPU continuously diagnoses the entire
system for correct operation and sample water flow. If an error occurs,
a message is posted to the message queue on the LCD screen.
The reagent is added by a single pulse of the reagent solenoid. When
the reagents require replacement, the
PRIME
button is pushed to bring
new reagents into the system. During
PRIME
the reagent solenoid is
pulsed several times to draw fluid from the two reagent bottles and
fill the tubes with new reagent. A complete
PRIME
takes less than
a minute.
The
SERVICE
button empties the cuvette, stops the flow of sample
water, and clears any errors. This provides a convenient way to replace
reagents and the measurement cuvette. If more extensive servicing
is performed, all power to the CLX-Ex2 should be removed. If the
CLX-Ex2 is to be turned off, it is recommended that the instrument be
placed in
SERVICE
mode before removing power. This ensures that
the cuvette is emptied and the flow is off.
3.2 The Air Sentinel Controller
The Air Sentinel controller governs the power to the instrument. For
normal instrument operation air needs to passes through fitting on
the base of the main enclosure and constantly purge any vapors that
could build-up. All GREEN lamps indicate a safe operating condition.
Only when this condition is reached will power be applied to the
measurement portion of the instrument.
Start-up
At start-up with the enclosure door open the Air Sentinel will display a
RED
PRESSURE
lamp on the visual indicator indicates that the system
is not pressurized and no power is sent to the instrument.
Closing the door with the air pressure applied, will cause the Air
Sentinel to display a GREEN
PRESSURE
lamp and a RED
TIMER
lamp. During this time the enclosure will undergo an automatic dilution
purge period of 5 minutes where at least 5 volumes of free space inside
the enclosure will be exchanged with clean purge air. No power is
applied to the rest of the instrument during this time.
After the enclosure air has gone through this purge period the Air
Sentinel will display all GREEN lamps. At this time the instrument is
deemed safe and power is applied to rest of the instrument.
Do not apply power to the Air Sentinel controller unless the area
has been properly tested and is known not to contain explosive
materials.
Bypass Key
Use of the bypass key disables the safety features and safe
operation of this instrument. It is imperative that this operation
be performed only after the area has been tested and known to
be safe.
If power is retained to the CLX-Ex2 after opening the enclosure
door while NOT in bypass mode, immediately close the
enclosure door and remove power to the instrument. Contact
HF scientific service department for further instruction.
Assuming the area has been tested and is known not to contain
explosive materials; a bypass service key is provided and can be
used. In this operating mode power will be retained to the rest of the
instrument even though the door is open. It is imperative that this mode
is only used after area is known to be safe. This operation disables the
safe operation of this instrument.
Upon turning the key, the indicator will display the RED
BYPASS &
TIMER
lamps. This is an indication of the maintenance mode. Power
is retained to the instrument. The key cannot be removed in this mode
of operation.
After maintenance has been performed the door should be closed
and latched. Air pressure will once again pressurize the enclosure. The
maintenance key can now be turned and removed and the key cover
replaced. The purge controller will display GREEN
PRESSURE &
BYPASS
lamps indicating a safe operation and power is kept applied
to the instrument. The
TIMER
lamp will remain RED for 5 minutes to
indicate a bypass operation had taken place.
WARNING
!