background image

Use

NEXT

to move the selection bar to the next setting.

Use

+/-

to alter the value of the highlighted setting:

l

The

contact dose

is the volume of fluid the pump will dispense when an external pulse

is received on input pin 2.

l

The

flow rate

will determine the time taken to complete each dose.

l

Contact memory

determines what the pump will do when pulses are received while a

dose is in progress. If set to “ignore” the pulses will be forgotten by the pump. If set to
“add” any pulses received while dosing will be queued up in memory and will be
dispensed when the current dose has finished.

If pulses are buffered in the memory then the pump will not stop between doses.

Once you have decided all your settings press

FINISH

and press

SAVE

to store.

Figure 23 - Electrical pulse specification

T

t

t

2

Vmax

V

0

V:

Voltage

T:

Time

Vmax:

Maximum voltage

t:

40 ms (min) to

1000 ms (max)

Vmin:

> 1.9 V

t

2:

> 1 s

m-qdos-allmodels-en-07

94

Содержание qdos 20

Страница 1: ...Watson Marlow qdos User Manual m qdos allmodels en 07 1...

Страница 2: ...ng 27 8 Pumphead materials of construction 28 8 1 Fluid path components 28 8 2 Materials of construction that have the potential of being wetted when the tube or fluid contacting element fails 29 9 Ac...

Страница 3: ...e PROFIBUS station address at the pump 68 20 2 PROFIBUS communication errors 70 20 3 PROFIBUS GSD file 72 20 4 User Parameter Data 74 20 5 PROFIBUS data exchange 75 20 6 Device related diagnostic data...

Страница 4: ...Drive maintenance 130 29 Pumphead Replacement qdos 30 131 29 1 Connecting interface tubing 136 30 Pumphead Replacement qdos 20 60 120 and CWT 139 30 1 Connecting interface tubing 143 31 Ordering info...

Страница 5: ...used BS EN61010 1 2010 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements EN61326 1 2013 Electrical equipment for measurement control...

Страница 6: ...ised of the possibility of such damages Watson Marlow shall not be liable for any loss damage or expense directly or indirectly related to or arising out of the use of its products including damage or...

Страница 7: ...s or accidental damage or neglect are excluded l Failure caused by electrical surge is excluded l Failure caused by incorrect or sub standard system wiring is excluded l Damage by chemical attack is e...

Страница 8: ...3 2 Inspection Check that all components are present Inspect components for damage in transit If anything is missing or damaged contact your distributor immediately 3 3 Components supplied Qdos 30 Cu...

Страница 9: ...rings x 2 qdos30 only for food applications l The designated power cable attached to the pump drive unit l Quick start manual l Product safety information booklet l Hydraulic connector pack The qdos1...

Страница 10: ...er care should be taken after storage to ensure that all parts function correctly Please observe the storage recommendations and use by dates which apply to ReNu pumpheads and tubing you may wish to b...

Страница 11: ...we will issue a Returns Authorisation Number We reserve the right to quarantine or refuse any equipment that is not displaying a Returns Authorisation Number Please complete a separate decontamination...

Страница 12: ...Caution hot surface Fundamental work with regard to lifting transportation installation starting up maintenance and repair should be performed by qualified personnel only The unit must be isolated fr...

Страница 13: ...ot during operation Do not take hold of the pump while it is running Allow the pump to cool after use before handling The drive unit must not be run without a pumphead fitted The pumphead should not b...

Страница 14: ...qdos 20 60 120 and CWT on page 139 To prevent dosing liquid entering the pump housing follow the recommended actions below To prevent damage to the pump and pumphead due to failure of the consumable...

Страница 15: ...ring the CONTROL SETTINGS via the MENU softkey ReNu PU Pumphead only when replacing a pumphead prior to the recommended lifespan of the tube or before a tube failure once the pump has been powered dow...

Страница 16: ...1 3 4 5 2 6 7 Figure 3 qdos 20 60 120 and CWT features 1 ReNu pumphead left hand mounted 2 Mounting plate 3 Pumphead locking latch 4 Outlet 5 Inlet 6 HMI not remote 7 Drive unit qdos 20 60 120 and CWT...

Страница 17: ...ure 4 qdos 20 60 120 and CWT with relay module features 1 ReNu pumphead left hand mounted 2 Mounting plate 3 Relay module 4 Pumphead locking latch 5 Outlet 6 Inlet 7 HMI not remote 8 Drive unit m qdos...

Страница 18: ...0 1 3 3 4 5 8 2 6 7 Figure 5 qdos 30 features 1 ReNu pumphead left hand mounted 2 Mounting plate 3 Pumphead retaining clamps 4 Outlet 5 Inlet 6 HMI not remote 7 Drive unit 8 Power m qdos allmodels en...

Страница 19: ...5 6 9 3 2 7 8 Figure 6 qdos 30 with relay module features 1 ReNu pumphead left hand mounted 2 Relay module 3 Mounting plate 4 Pumphead retaining clamps 5 Outlet 6 Inlet 7 HMI not remote 8 Drive unit 9...

Страница 20: ...ils product reference number serial number and model details 190VA 100 240VAC 50 60Hz 0M0 223L GLU Iss 1 qdos Manual 30l hr 7bar 190gpd 100psi Sant PFPE Serial Number 1 2 3 4 Figure 7 Nameplate 1 Manu...

Страница 21: ...50 60 Hz AC Power consumption 190 VA DC Supply voltage 12 24 VDC power option 12 24 V DC DC Power consumption 12 24 VDC power option 150 W Installation category overvoltage category II 10 of nominal...

Страница 22: ...Standards UL 61010A 1 UL CSA 61010 1 CAN CSA C22 2 No 61010 1 IEC 61010 1 Radiated emissions FCC 47CFR Part 15 NEMA 4X to NEMA 250 NSF61 Not ReNu PU pumpheads EC 1935 2004 and EU regulation 10 2011 c...

Страница 23: ...A 250 NSF61 EC 1935 2004 and EU regulation 10 2011 certification FDA regulation 21CFR parts 170 199 6 4 Dimensions A B C D F G H I E Figure 8 Dimensions Dimension qdos20 qdos30 qdos60 qdos120 qdos CWT...

Страница 24: ...266 mm 10 5 266 mm 10 5 290 9 mm 11 5 E Optional relay modules 43 mm 1 7 43 mm 1 7 43 mm 1 7 43 mm 1 7 43 mm 1 7 F 173 mm 6 8 173 mm 6 8 173 mm 6 8 173 mm 6 8 173 mm 6 8 G 40 mm 1 6 40 mm 1 6 40 mm 1...

Страница 25: ...iversal 4 6 10 lb 2 oz 5 7 12 lb 9 oz 6 8 15 lb 0 oz Universal 4 6 10 lb 2 oz 5 7 12 lb 9 oz 6 8 15 lb 0 oz PROFIBUS 4 6 10 lb 2 oz 5 7 12 lb 9 oz 6 8 15 lb 0 oz Universal 24 V relay 4 8 10 lb 9 oz 5...

Страница 26: ...V relay 4 3 9 lb 8 oz 5 25 11 lb 9 oz Universal 110V relay 4 3 9 lb 8 oz 5 25 11 lb 9 oz qdosCWT Table 3 Weights qdos CWT Model Drive Drive with pumphead kg lb kg lb Manual 4 1 9 lb 6 3 13 lb 14 oz Re...

Страница 27: ...ant contained in the ReNu Pumphead For guidance refer to www qdospumps com Table 4 Materials of construction for external casing Component Pump Variant qdos 20 qdos 30 qdos 60 qdos 120 qdos CWT Keypad...

Страница 28: ...www qdospumps com 8 1 Fluid path components Table 5 Fluid path component materials of construction Component qdos 20 qdos 30 qdos 60 qdos 120 qdos CWT Tube or fluid contacting element SEBS TPU SEBS Sa...

Страница 29: ...umphead enclosure 30 GF PPE PS PC PP 316 stainless steel 40 GF PPS 20 GF PP PC PA6 316 stainless steel 30 GF PPE PS PC PP 316 stainless steel 40 GF PPS Pumphead enclosure seals NBR NBR NBR EPDM NBR Ro...

Страница 30: ...Polyamide PA6 Nylon 6 PC Polycarbonate PEEK Polyether ether ketone PFPE Perfluoropolyether POM Polyoxymethylene PP Polypropylene PPE Polyphenyl ether PPS Polyphenylene sulfide PS Polystyrene PVCU Pol...

Страница 31: ...to avoid the sudden release of pressurised fluid in the event of a pumphead failure This shall be fitted immediately after the discharge of the pump qdos is a positive displacement pump therefore it...

Страница 32: ...suitably rated to handle the predicted pipeline pressure Avoid pipe reducers and lengths of smaller bore tubing than the pumphead section particularly in pipelines on the suction side Any valves in t...

Страница 33: ...tem and to allow for the safe disposal of the pumphead the ignore function can be used to operate the pump prior to changing the pumphead Only applicable for pumps manufactured up until October 2019 D...

Страница 34: ...s will continue to operate when there is gas present in the suction line and will maintain prime in these conditions The pump can be run dry however flow rate and pumphead life will be affected m qdos...

Страница 35: ...e disassembled Each pump is supplied with a power cable The gland at the pump end of the cable is IP66 rated The mains plug at the opposite end of the cable is NOT IP66 rated Ensure that all power sup...

Страница 36: ...of the pump should not be disassembled Installation instructions It is highly recommended that an isolating switch is provided between the power source and the pump The cable is fitted with M8 ring te...

Страница 37: ...n O ring seals 2 off with the EPDM O ring seals supplied Fit the pumphead to the drive Ensure the pumphead ports are securely fitted to the interface tubing Ensure proper connection has been made to a...

Страница 38: ...he user s responsibility to ensure the safe and reliable operation of the pump under remote and automatic control Never apply mains power to the five pole M12 connectors Apply the correct signals to t...

Страница 39: ...nimum pulse length referenced to pin 4 Alternatively connect pin 5 of the output to this pin via normally open switch White 3 4 20 mA 250 input impedance 40 mA max current 250 load resistance 40 mA ma...

Страница 40: ...40 mS maximum duration 1000 mS This input is used to trigger a user defined dose size Refer to the contact mode section Speed analog input The speed of this pump can be controlled remotely by a curre...

Страница 41: ...connected via a NO switch to input pin 1 or 2 to power the inputs Grey 2k2 Typical 1k2 Minimum 1 4 1 2 3 4 5 PLC 24 VDC Input Common Note Diagram depicts Run Status output Figure 14 Example wiring for...

Страница 42: ...nt analog signal within the range 4 20 mA into 250 impedance is available between pin three and pin four of the output connector The current is fixed and directly proportional to the pumphead rotation...

Страница 43: ...ing the relay module cover The module cover is secured to the side of the drive unit by four M3x10 Pozidriv pan head stainless steel screws Remove the four screws from the module cover leaving the top...

Страница 44: ...s are provided with the pump Recommended control cable conductors for the terminal blocks metric 0 05 1 31 mm2 USA 30 16 AWG stranded and solid Cable circular Max min outside diameter to ensure a seal...

Страница 45: ...y be convenient to strip the outer sheath before passing the cable s conductors through the gland However to ensure a watertight seal the cable must have an undamaged sheath within the gland when it i...

Страница 46: ...ensure a watertight seal Alternatively tighten the gland by hand until it is finger tight and use a spanner to tighten it further by one half a turn Ensure that unused openings on the module are seal...

Страница 47: ...Universal relay variant has two terminal blocks with two output options Figure 17 Universal 4 relay PCB Never apply mains power to the 4 20 mA input 4 20 mA output or stop contact terminals Apply the...

Страница 48: ...rol settings menu 3 N C 2 C 1 N O 3 2 1 Table 9 Run status output J2 RLY2 Connect the output device to the C common terminal of the relay connector and either the N C normally closed or N O normally o...

Страница 49: ...t terminal and the 5 V terminal of the Run Stop I P connector J8 Alternatively a 5 V 24 V logic input may be applied to the Stop Contact terminal ground to the GND terminal of the adjacent 4 20 mA I P...

Страница 50: ...to the I P terminal of the Analog connector J4 Ground to the GND connector of the same terminal In Analog mode the pump set speed will be proportional or inversely proportional to the analog input 4...

Страница 51: ...ed for earth shielding of cables Earth can be connected to the terminal There are also two spring clamp terminals for additional earth connection Keep 4 20 mA and low voltage signals separate from mai...

Страница 52: ...ing resistor The M12 socket provided for PROFIBUS installation is IP66 rated To maintain an IP66 rated system the PROFIBUS cable T adaptors and terminating resistors used must be fitted with IP66 rate...

Страница 53: ...V supply for terminating resistors 2 RxD TxD N Data line minus A line 3 DGND Data ground 4 RxD TxD P Data line plus B line 5 Shield Ground connection Note If the pump is the last bus device connected...

Страница 54: ...nual PROFIBUS Universal and Universal only Power up the pump The pump displays the start up screen with the Watson Marlow Pumps logo for three seconds Selecting your chosen display language Use the ke...

Страница 55: ...guage Choose REJECT to return to the language choice screen The pump is preset with operational parameters as shown in the table below Confirm which ReNu pumphead is installed qdos20 and qdos60 only U...

Страница 56: ...You have selected PU Choose REJECT to return to the pumphead select screen The pump is preset with operational parameters as shown in the table below m qdos allmodels en 07 56...

Страница 57: ...unit ml min Backlight 30 minutes Pump tag WATSON MARLOW Auto restart Off This then proceeds to the home screen The pump is now ready to operate according to the defaults listed above Note The display...

Страница 58: ...in place when the pump was switched off last Check that the pump is set to operate as you require it The pump is now ready to operate All operating parameters may be changed by means of key presses s...

Страница 59: ...ching on Remote When power is supplied to the pump all the LED icons will illuminate for three seconds After this period the pump will operate in accordance with the inputs received m qdos allmodels e...

Страница 60: ...portionally to the analog signal provided The default figures are 4 1 mA 0 rpm 19 8 mA 125 rpm To stop the pump remotely apply a signal minimum 5 V maximum 24 V to input pin 1 To run the pump in rever...

Страница 61: ...nu 5 RIGHT HAND FUNCTION When pressed will perform the function displayed on the screen directly above the relevant function key 6 LEFT HAND FUNCTION When pressed will perform the function displayed o...

Страница 62: ...ing the START key in manual mode or by selecting Analogue mode In this state the pump will respond to a change in state of the start stop input and may start automatically when a control signal is rec...

Страница 63: ...odes The available modes are l Manual default l Flow calibration l Analog 4 20mA Universal and Universal only l Contact mode Universal only l PROFIBUS PROFIBUS only l Fluid recovery l CANCEL Use SELEC...

Страница 64: ...n normal operation the direction of flow is into the bottom port of the pumphead and out of the top port The symbol is a warning that the pump could remotely activate at any time It will always be dis...

Страница 65: ...DECREASING FLOWRATE Using the keys will increase or decrease the flowrate Decreasing flowrate l A single key press will decrease flowrate by the least significant digit of the chosen flowrate unit l R...

Страница 66: ...ld down the key for flowrate scrolling Max 100 function Manual mode only l Press and hold the MAX key to run at maximum flow l Release the key to stop the pump l The volume dispensed and time elapsed...

Страница 67: ...n only be set from the pump The user can set the station address within this mode Select MODE Using the keys scroll to PROFIBUS and press SELECT If PROFIBUS is not enabled the screen below will prompt...

Страница 68: ...Assigning the PROFIBUS station address at the pump The station address can only be set from PROFIBUS settings The station address cannot be automatically assigned by the master Select MODE Using the...

Страница 69: ...keys alter the station address in the range from 1 to 125 126 is the default station address Press FINISH to set the station address or NEXT to enable disable PROFIBUS communication m qdos allmodels...

Страница 70: ...een below is displayed the P indicates data exchange is happening This screen will only be displayed after successful implementation of Master Slave communications which always follow the sequence des...

Страница 71: ...e following screen will be displayed The first red dot corresponds to the stage at which the error occurred and subsequent stages will indicate a red dot because the communication sequence halted befo...

Страница 72: ...er on interrogation The GSD file file name WAMA0E7D GSD can bedownloaded from the Watson Marlow website and installed or typed into a PROFIBUS master directly from this manual using a GSD editor progr...

Страница 73: ...supp 1 MaxTsdr_9 6 60 MaxTsdr_19 2 60 MaxTsdr_45 45 60 MaxTsdr_93 75 60 MaxTsdr_187 5 60 MaxTsdr_500 100 MaxTsdr_1 5M 150 MaxTsdr_3M 250 MaxTsdr_6M 450 MaxTsdr_12M 800 Slave_Family 0 Implementation_Ty...

Страница 74: ...Byte 7 Fail Safe 8 bit Byte 8 Fail Safe Speed Low byte of 16 bit unsigned 8 bit Byte 9 Fail Safe Speed High byte of 16 bit unsigned Set Min Max Speeds The Min Max Speed parameters are used to set the...

Страница 75: ...US network literature for further information Table 23 PROFIBUS data exchange Default address 126 PROFIBUS Ident 0x0E7D GSD File WAMA0E7D GSD Config 0x62 0x5D 3 words out 14 words in User parameter by...

Страница 76: ...Cyclic Data Read from pump to Master Table 26 Cyclic data read from pump to Master 16 bit Byte 1 2 Status Word 16 bit Byte 3 4 Pumphead Measured Speed unsigned 16 bit Byte 5 6 Hours Run 16 bit Byte 1...

Страница 77: ...s counts down from FF for each complete motor revolution Reset this counter to FF by using bit 2 of the control word The motor relates to the motor inside the pump before the gearbox ratio The number...

Страница 78: ...Software Version Main CPU 32 bit Byte 14 15 16 17 Software Version HMI CPU 32 bit Byte 18 19 20 21 Software Version Flash 32 bit Byte 22 23 24 25 Software version PROFIBUS CPU 20 7 Channel related dia...

Страница 79: ...voltage 0xA3 Over voltage 0xA3 Over voltage Motor stall 0xA4 Overload Over temp 0xA5 Over temp 0xA5 Over temp Tacho fault 0xB1 Device related 0x11 Leak detected 0xB2 Device related 0x12 Fluid level al...

Страница 80: ...and Universal only This pump displays flow rate in ml min Setting the flow calibration Select MODE Using the keys scroll to Flow calibration and press SELECT Using the keys enter the maximum flow rat...

Страница 81: ...Press START to begin pumping a volume of fluid for calibration Press STOP to stop pumping fluid for the calibration m qdos allmodels en 07 81...

Страница 82: ...keys enter the actual volume of fluid pumped To accept the new calibration press ACCEPT or RECALIBRATE to repeat the procedure Press HOME or MODE to abort The pump is now calibrated m qdos allmodels...

Страница 83: ...liamp signal and the flow rate is determined by configuring the two points A and B as shown in the graph below The rate of flow can be proportional or inversely proportional to the analogue milliamp i...

Страница 84: ...Using the keys scroll to Analog 4 20 mA and press SELECT The current signal being received by the pump is displayed for information only on the HOME screen Pressing the INFO function key will display...

Страница 85: ...20 mA profile by using a multiplication factor Press from the HOME screen to access the scaling factor Use the keys to enter a multiplication factor A figure of 1 00 will not alter the 4 20 mA profil...

Страница 86: ...s ACCEPT to confirm the new 4 20 mA PROFILE figures This will not alter the stored A and B points the multiplication factor will re scale the 4 20 mA profile To re set the original flow rates re set t...

Страница 87: ...n all modes The speed limit cannot exceed the high flow rate set point B The speed limit function takes precedence over the scaling factor For example if the qdos20 4 20 mA profile is 0 flow at 4 mA t...

Страница 88: ...gnals must be within range If the signal sent is out of range you will not be able to set the signal input value and progress to the next step in the process Select MODE Using the keys scroll to Analo...

Страница 89: ...input to the pump or enter the current value using the keys ACCEPT appears when high 4 20 mA signal is within tolerance limits Press ACCEPT to accept the high signal input or CANCEL to return to the...

Страница 90: ...ng the keys scroll to the desired flow rate Select SET FLOW or press BACK to return to the previous screen Setting a low signal Send the low signal input to the pump or enter the current value using t...

Страница 91: ...al is less than 1 5 mA the following error message will be displayed ACCEPT appears when low mA signal is within tolerance limits Press ACCEPT to accept the low signal input or CANCEL to return to the...

Страница 92: ...ng the keys scroll to the desired flow rate Select SET FLOW This proceeds to the screen confirming calibration is complete Select CONTINUE to start in proportional mode or MANUAL to continue in manual...

Страница 93: ...is received The dose volume is a user defined value between 0 1 ml and 999 l 23 1 Contact settings To setup contact mode firstly you will need to define the settings To do this press the MODE key mov...

Страница 94: ...re received while a dose is in progress If set to ignore the pulses will be forgotten by the pump If set to add any pulses received while dosing will be queued up in memory and will be dispensed when...

Страница 95: ...ode The contact mode home screen will be displayed The home screen displays the contact dose the flow rate and the dose time remaining when a dose is in progress The dose time is only displayed on scr...

Страница 96: ...intervals We recommend examining your process to identify whether a 4 20 mA signal can be used rather than a pulse Where technology does not allow a 4 20 mA signal we recommend using a signal converte...

Страница 97: ...t your system design permits reverse flow If unidirectional valves are installed then reverse flow will not function and the pump will build up excessive pressure within the pipework Press and hold RE...

Страница 98: ...onality To enable this position the selection bar over fluid recovery in the mode menu and press the SETTINGS function key Select ENABLE to turn this functionality on The reverse process can be used t...

Страница 99: ...4 volts to input pin 1 2 Apply 5 24 volt to pin 5 of the pump input 3 Remove the remote stop signal 4 Apply 4 20 mA to the analog input This will cause the pump to run in reverse at a speed proportion...

Страница 100: ...HOME screens or INFO screens For example Manual HOME screen Manual INFO screen This will display the main menu as shown below Use the keys to move the selection bar between the available options Press...

Страница 101: ...ervoir will reduce the bar will track the reducing volume The pump can be set up to output an alarm when a defined level of fluid is reached This will warn the operator to change the fluid supply barr...

Страница 102: ...changed by pressing the US GALLONS or LITRES key the key name will change depending upon the unit selected To configure the level monitor select this option from the menu Enter the maximum level of y...

Страница 103: ...ss SELECT to return to the fluid level monitor menu If you need to adjust the volume of fluid in the tank for example when re filling then press SELECT when the bar highlights the Adjust level option...

Страница 104: ...settings Manual PROFIBUS Universal and Universal only Security settings can be changed by selecting SECURITY SETTINGS from the Main menu Auto keypad lock Press ENABLE DISABLE to switch on off the Auto...

Страница 105: ...unlock the keypad press the two unlock keys together The padlock icon will appear on the operating mode home screen to show that keypad lock is activated Note that the STOP and BACKLIGHT keys will alw...

Страница 106: ...ny change of operating mode settings or entry to the menu Once a correct PIN has been entered all settings can be changed PIN protection will turn back on automatically after 20 seconds of no keypad a...

Страница 107: ...Pressing the HOME or MODE key at any time before you confirm your PIN will abort the process If you enter an incorrect PIN it will display the following screen There is an override feature should you...

Страница 108: ...it would return to the same operating mode and continue to run at a proportional speed to the analog input If the pump was running in manual mode prior to power loss it would return to the same opera...

Страница 109: ...a power cycle auto restart applies to manual mode network mode and memodose Flow units The current chosen flow unit is displayed on the right hand side of the screen To change flow units move the sele...

Страница 110: ...through the available characters for each digit The available characters are 0 9 A Z and SPACE Press NEXT to move onto the next character or PREVIOUS to move back to the previous character Press FINIS...

Страница 111: ...k to the last character Press FINISH to save the entry and return to the general settings menu Restore defaults To restore the factory default settings select restore defaults from the general setting...

Страница 112: ...efore changing the language Use the keys to move the selection bar to your required language Press SELECT to confirm Your selected language will now be displayed on screen Press CONFIRM to continue al...

Страница 113: ...L SETTINGS from the main menu to access the sub menu shown below Use keys to move the selection bar Press SELECT to choose the required function Speed limit The maximum speed the qdos30 qdos60 or qdos...

Страница 114: ...g modes Figure 24 The effect of a 75 rpm speed limit on user defined 4 20 mA response profiles mA rpm mA rpm Calibrated 4 20 mA user_max_flow recalibrated Figure 25 The effect of a 30 rpm speed limit...

Страница 115: ...speed limit to 125 rpm when the power is turned off to the pump If you require this functionality on a software version lower than MKS 2 0 please use the 4 20 mA calibration method described in Secti...

Страница 116: ...ome screen Invert alarm logic Universal model Select invert alarm logic from the control settings menu Select ENABLE to invert the alarm output The default setting is high for healthy low for alarm It...

Страница 117: ...Use and SELECT to choose which output to configure Use and SELECT to choose which pump status you require for the chosen output The tick symbol indicates the current setting m qdos allmodels en 07 117...

Страница 118: ...the chosen output Press SELECT to program the output or Exit to cancel 4 20 mA Output Universal model only Select 4 20 mA to configure the pumps 4 20 mA output response Use and SELECT to choose the r...

Страница 119: ...to the same range as the 4 20 mA input If the 4 20 mA input has been scaled to provide 4 mA 0 rpm and 20 mA 20 rpm then an input of 12 mA will result in a set speed of 10 rpm and an output of 12 mA T...

Страница 120: ...e the input setting A low stop input is recommended as the pump will be stopped in the event of any loss of input signal To disable remote automatic control at the pump select CONFIGURE DISABLE INPUT...

Страница 121: ...om to Note This will only disable the remote stop functionality in manual mode The remote stop cannot be disabled in analogue mode Note You must manually change back to remote automatic control manual...

Страница 122: ...ion qdos20 and qdos60 only To change the pumphead selection from one material to another or to confirm that the pumphead has been replaced early use and press SELECT to make the selection m qdos allmo...

Страница 123: ...24 6 Help Manual PROFIBUS Universal and Universal only Select Help from the main menu to access the help screens m qdos allmodels en 07 123...

Страница 124: ...ach error state is provided in the table below Figure 26 Status LEDs Status 4 20 mA Running Remote stop Change pumphead 4 20 mA signal Power on On 4 20 mA in range On On 4 20 mA high On Flash 4 20 mA...

Страница 125: ...g if one is present If the pump runs but there is little or no flow make the following checks l Check that fluid is supplied to the pump l Check for any kinks or blockages in the lines l Check that an...

Страница 126: ...tly orientated with the arrow pointing upwards If the message constantly repeats after several pumphead installations then there may be a leak detection sensor failure If the leak detector is faulty p...

Страница 127: ...t the nature of an external condition They do not flash Table 35 Error codes Error code Error condition Suggested action Er 0 FRAM write error Attempt to reset by switching power OFF ON Or seek suppor...

Страница 128: ...f range Check analog control signal range Trim signal as required Or seek support Er 21 Over signal Reduce the analog control signal Er 50 Communication error Attempt to reset by switching power OFF O...

Страница 129: ...27 Technical support Support is available from your local Watson Marlow representative Web www wmftg com m qdos allmodels en 07 129...

Страница 130: ...Drive maintenance There are no user serviceable parts inside the pump The unit should be returned to Watson Marlow for service See Information for returning pumps on page 11 m qdos allmodels en 07 13...

Страница 131: ...for contaminated items and health and safety procedures Always isolate the pump from the mains power supply before changing the pumphead suction or discharge lines It is only possible to fit the pumph...

Страница 132: ...tion replace the existing FKM Viton O ring seals 2 off with the EPDM O ring seals supplied It is the responsibility of the end user to ensure that the pumphead and seals comply with the migration and...

Страница 133: ...the mains power supply 4 Ensure that protective clothing and eye protection are worn if hazardous products have been pumped 5 Remove input and output connections from the pumphead protecting the pump...

Страница 134: ...and rotate it in an anti clockwise direction by approximately 15 8 Remove the pumphead from the pump housing 9 Safely dispose of the used pumphead according to your own health and safety regulations...

Страница 135: ...e pump drive shaft and slide into position on the pump housing 4 Rotate the pumphead in a clockwise direction by approximately 15 to engage the retaining clamps 5 Tighten the retaining simultaneously...

Страница 136: ...nsure that the correct seal material FKM Viton 1a or EPDM 1b has been selected and are correctly fitted in the head ports 1 and that the seal material and connector material is compatible with pumped...

Страница 137: ...ure that the product is installed correctly according to the manufacturers instructions 1 Select the correct compression fittings for the size of tube using the markings on the sprue and detach both t...

Страница 138: ...X X X X X X X 1 4 3 2 2a 3a 3b 4a 4b 4c 4d 4e 4f M24x2mm 1 SAFETY OVERFLOW 5x8 6 3x11 5 9x12 5x8 6 3x11 5 9x12 10x16 10x16 Figure 29 ReNu 30 safety overflow A Safety overflow m qdos allmodels en 07 13...

Страница 139: ...gulations for contaminated items and health and safety procedures Always isolate the pump from the mains power supply before changing the pumphead suction or discharge lines Always operate the ReNu pu...

Страница 140: ...ounted pumphead Replacing the right hand mounted pumphead is the identical procedure Removing the pumphead 1 Drain down 2 Ensure that there is no pressure in the pipeline 3 Isolate the pump from the m...

Страница 141: ...the pressure switch back to transport position not required for ReNu CWT Transport position P 1 bar 15 psi 8 Safely dispose of the used pumphead according to your own health and safety regulations Ta...

Страница 142: ...bar 15 psi 3 Align the new pumphead with the pump drive shaft and slide into position on the pump housing 4 Rotate the pumphead in an anti clockwise direction by approximately 15 to engage the retain...

Страница 143: ...eals and connector material is compatible with pumped fluid the moulded seals supplied with the pumphead are the same material as the pumphead tube ReNu PU pumpheads are supplied with TPU seals ReNu S...

Страница 144: ...tting and tighten onto the pumphead 3a and 3b 3 When attaching the mating thread restrain the connector using a 14 mm spanner for the 1 4 BSPT 3a 9 16 spanner for the 1 4 NPT 3b 1 2 spanner for the 1...

Страница 145: ...may be necessary to widen the end of the tube 6 Whilst continuing to hold the tube against the back face of the cone slide the compression ring and user connection collar back down the tube and tighte...

Страница 146: ...6 3 x 11 5 9 x 12 10 x 16 10 x 16 SAFETY OVERFLOW Figure 33 ReNu 20 60 120 and CWT safety overflow A Safety overflow The ReNu CWT pumphead will vary in appearance from the ReNu 20 60 and 120 pumphead...

Страница 147: ...a V 12 24V DC Fluid path material 2 Santoprene 5 PU 7 EPDM 8 SEBS The pumphead side location is required when ordering The left right perspective assumes the user is looking at the front of the pump T...

Страница 148: ...0M3 2800 PFP qdos60 0M3 3800 PFP ReNu PU pumphead PFPE Lubricant qdos20 PU 0M3 1500 PFP ReNu CWT pumphead PFPE lubricant qdos CWT 0M3 5700 PFP Hydraulic connection pack polypropylene compression fitt...

Страница 149: ...30 pumpheads Note Remove the standard seal before fitting this connector 0M9 401H P04 Hydraulic connection pack polypropylene 1 2 hose barb 0M9 401H P05 Hydraulic connection pack PVDF threaded fitting...

Страница 150: ...e t h y l e n e Interface tubing polyethylene 9x12 mm 2 m 6 5 ft length 0M9 2222 E9C P o l y e t h y l e n e Interface tubing polyethylene 5x8 mm 2 m 6 5 ft length 0M9 2222 E58 P o l y e t h y l e n e...

Страница 151: ...221R D00 ReNu 20 ReNu 60 ReNu 120 and ReNu CWT Santoprene Insert 0M9 001R M00 ReNu 20 ReNu 60 SEBS fluid port seal 0M9 001R B00 ReNu 20 ReNu 60 PU fluid port seal 0M9 001R A00 Qdos 30 pumphead clamp...

Страница 152: ...ontinuously at discharge pressures of up to 7 bar 100 psi qdos20 PU can be operated continuously at discharge pressures of up to 4 bar 60 psi qdos CWT can be operated continuously at discharge pressur...

Страница 153: ...ts Units Comment Min Nom Max Inrush current 17 A No Load Inrush current duration 20 mS Efficiency ring terminals 87 91 95 100W 10 12 24V Typical qdos pump power required 5 120 W qdos 20 30 60 120 CWT...

Страница 154: ...0 12 14 16 18 20 22 24 26 28 30 32 400 600 800 1000 1200 1400 1600 1800 2000 2200 15 10 5 0 0 20 40 60 80 100 34 Santoprene or SEBS Tubing Flowrate ml min 1 Flowrate ml min 1 Flowrate US GPH Flowrate...

Страница 155: ...gauge pressure psi Suction gauge pressure bar with 1 5 bar discharge gauge pressure Discharge gauge pressure bar with 0 bar suction gauge pressure 200 300 100 1 0 5 0 0 1 2 3 4 5 6 7 2 4 6 8 10 12 400...

Страница 156: ...33 Trademarks Watson Marlow qdos qdos20 qdos30 qdos60 qdos120 qdos CWT and ReNu are trademarks of Watson Marlow Limited m qdos allmodels en 07 156...

Страница 157: ...nual Updated to include qdos CWT models Updated to include 4 relay models First published 02 2020 m qdos en 06 Minor updates following review First published 02 2021 m qdos en 07 Updated for conforman...

Страница 158: ...stop input J1A 110 V logic 50 Table 15 Speed analog input J4 50 Table 16 Speed analogue output J3 Universal only 50 Table 17 Earth shielding terminals 51 Table 18 Max length of type A bus cable m 52 T...

Страница 159: ...le wiring for a pull up resistor 41 Figure 15 Example wiring for an external relay 41 Figure 16 Optional output lead 42 Figure 17 Universal 4 relay PCB 47 Figure 18 PROFIBUS connector location 52 Figu...

Страница 160: ...Figure 33 ReNu 20 60 120 and CWT safety overflow 146 Figure 34 Left hand mounted pumphead 147 m qdos allmodels en 07 160...

Отзывы: