Pump operation
Note:
Sections "Keypad functions (Manual, PROFIBUS, Universal and Un only)"
below up to and including "Help (Manual, PROFIBUS, Universal and Un only)" on
page 123 apply to Manual, PROFIBUS, Universal and Un only. It is only possible to
control the Remote pump via the input and output, (I/O) provided.
18.1
Pump operation (Remote pump)
The remote pump will operate proportionally to the analog signal provided. The default
figures are 4.1 mA = 0 rpm, 19.8 mA = 125 rpm.
To stop the pump remotely apply a signal, minimum 5 V, maximum 24 V, to input pin 1. To
run the pump in reverse, apply a signal, minimum 5 V, maximum 24 V to input pin 5.
18.2
Pump operation (Manual, PROFIBUS, Universal and
Un only)
Keypad functions (Manual, PROFIBUS, Universal and Un only)
2
3
4
7
8
9
10
1
6
5
Figure 20 - Keypad functions
1.
Colour TFT display
2.
Backlight
3.
+/- Keys
4.
Mode
5.
Right hand function
6.
Left hand function
7.
Max/Priming
8.
Stop
9.
Start
10.
Home