Watson-Marlow 720DuN User Manual
34
Calibration dose
The pump offers to run a calibration dose. Use the
UP
and
DOWN
keys to
choose
Yes
or
No
. Press
ENTER
to confirm. Examples are shown here.
If
No
is selected after a tube selection has been made (see
Tube
above), the
pump briefly displays a confirmation screen showing current head, tube and
flow settings, and redisplays the main menu.
If
Calibration dose
is selected or if
Yes
is selected after a tube selection has
been made (see
Tube
above), the pump displays the speed and direction at
which it was last running in manual mode or to which it has just been set and
invites the user to press
START
.
Note
: it is possible at this point to change the direction of rotation using the
DIRECTION (SHIFT, 1)
key, and to change the speed of rotation using the
numeric keys (followed by
ENTER
), or the
UP
and
DOWN
keys, within the
minimum and maximum limits that have been set. See 18.13
Set maximum
allowed speed
and 18.14
Set minimum allowed speed
.
Put a measuring container at the pump outlet. Press
START
. The pump runs
for 4 minutes, displaying an information screen for 15 seconds and a further
information screen for the rest of the 4 minutes. You may stop the calibration
dose at any time with the
STOP
key - but allow the pump to run as long as
possible to obtain the most accurate calibration. A minimum of 15 seconds is
recommended.