Watson-Marlow 323Dz User Manual
34
15.4.9 Proceed
If you wish to begin the dosing session, use the
UP
or
DOWN
keys to highlight
Proceed
and press
RETURN
. The pump offers the user the opportunity to
prime the pump. To do so,
press and hold
the
START
key. The pump operates
continuously at maximum speed until the
START
key is released. When priming
is complete, or if no priming is required, press
RETURN
. The pump displays the
dose size and the number of doses and waits for the
START
key to be pressed.
If you wish to review or change the parameters for the next batch, press
STOP
twice and use the
UP
and
DOWN
keys to move back up through the list of pa-
rameters.
15.4.10 Single dose dispensing
When
START
is pressed, the pump performs a dosing batch according to the pa-
rameters set.
If the batch is set to perform one dose, as in the example shown here, the pump
stops when that dose has been dispensed and waits for a further press on the
START
key. As each dose is dispensed, the display increments, recording the number of
doses dispensed so far.
Note
: To dispense a batch of one dose and set the pump to count the doses/batches,
check that the dose interval is set to zero seconds. If an interval is set, the pump will
still dispense batches of one dose, but will not increment its display to show how
many doses/batches have been dispensed.