background image

Watson Industries, Inc. 

 

ADS-C232-3AD Rev A  10/25/2017 

 

18 

 
3)  Change the Analog Bank Range to a new setting. 
 

To change the Bank Angle to Scale Factor 6.0°/V ,  type ‘1’. 
To change the Bank Angle to Scale Factor 9.0°/V ,  type ‘2’. 
To change the Bank Angle to Scale Factor 12°/V ,  type ‘3’. 
To change the Bank Angle to Scale Factor 18°/V ,  type ‘4’. 
To change the Bank Angle to Scale Factor 36°/V ,  type ‘5’. 

 

 

 
4)  To update the scale factor for the analog Elevation output, type ‘4’ at the main menu. 
This will bring up the menu:  
 

 
5)  To change the Elevation Angle setting. type ‘2’ at this menu. 
This will bring up the menu: 

 

6)  Change the Analog Elevation Range to a new setting. 

 

To change the Elevation Angle to Scale Factor 6.0°/V ,  type ‘1’. 
To change the Elevation Angle to Scale Factor 9.0°/V ,  type ‘2’. 
To change the Elevation Angle to Scale Factor 12°/V ,  type ‘3’. 
To change the Elevation Angle to Scale Factor 18°/V ,  type ‘4’. 
To change the Elevation Angle to Scale Factor 36°/V ,  type ‘5’. 

  
 
 

          THE CURRENT SETTING IN DEG/V IS:  6.0 
TYPE IN THE NUMBER OF YOUR SELECTION (OR 'Q' TO QUIT): 
          1 = BANK ANGLE SCALE FACTOR 6.0 DEG/V 
          2 = BANK ANGLE SCALE FACTOR 9.0 DEG/V 
          3 = BANK ANGLE SCALE FACTOR 12 DEG/V 
          4 = BANK ANGLE SCALE FACTOR 18 DEG/V 
          5 = BANK ANGLE SCALE FACTOR 36 DEG/V 
 

SELECT ATTITUDE FOR SCALE FACTOR UPDATE (OR 'Q' TO QUIT): 
          1 = ANALOG BANK ANGLE  
          2 = ANALOG ELEVATION ANGLE  
 

         THE CURRENT SETTING IN DEG/V IS:  6.0 
TYPE IN THE NUMBER OF YOUR SELECTION (OR 'Q' TO QUIT): 
         1 = ELEVATION ANGLE SCALE FACTOR 6.0 DEG/V 
         2 = ELEVATION ANGLE SCALE FACTOR 9.0 DEG/V 
         3 = ELEVATION ANGLE SCALE FACTOR 12 DEG/V 
         4 = ELEVATION ANGLE SCALE FACTOR 18 DEG/V 
         5 = ELEVATION ANGLE SCALE FACTOR 36 DEG/V 
 

Содержание ADS-C232-3AD

Страница 1: ...232 3AD Rev A 10 25 2017 1 Dual Axis Vertical Reference System Owner s Manual PART NUMBER ADS C232 3AD WATSON INDUSTRIES INC 3035 MELBY STREET EAU CLAIRE WI 54703 Phone 715 839 0628 Fax 715 839 8248 Email support watson gyro com ...

Страница 2: ...e Main Menu 16 Appendix B 1 Adjust Time Constants 16 Appendix B 2 Set Output Channels 16 Appendix B 3 List Output Channels 17 Appendix B 4 Select Analog Attitude Full Scale Range 17 Appendix B 5 Set Baud Rate 19 Appendix C 20 Watson Industries prides itself on solving customer problems and serving their needs in a timely fashion This manual is intended to facilitate this goal and to provide writte...

Страница 3: ...placement for mechanical gyros which are often more expensive and which cannot provide the reliability of a solid state sensor Theory of operation is described with more detail later in this manual The specifications size and pinouts are also described later in this document The Watson Vertical References provide for two types of analog output voltages One is proportional to angular rate and the o...

Страница 4: ...induced from other loads on the circuit Theory of Operation In order to obtain an accurate signal in both the short term and in the long term the Vertical Reference relies upon a Watson Industries angular rate sensor ARS combined with solid state accelerometers The position output is generated by integrating the angular rate output This signal is compared with the output from the accelerometer and...

Страница 5: ...ation data In most cases these algorithms provide for very stable and accurate output signals however the user should be aware that exceeding the normal ranges of the sensors would affect accuracy For example while custom ADS units may have differing ranges the rate gyros in a typical ADS have a maximum rate of 100 s Care should be taken not to exceed this rate because the closed loop system will ...

Страница 6: ...The contents of a typical string are formed as follows See Appendix B for information on how to change the data string 1 A single letter and a space used to indicate the start of the data string The letter I indicates the start of an inertial data string The letter R indicates the start of a Reference data string If the letter is in lower case i or r an error over range condition is indicated see ...

Страница 7: ... exists The system will continue to operate in an extended time constant mode with a low level of error accumulation until the condition is cleared Occasional blips of this condition are expected with no detectable affect on the resulting data The other output format available is a binary format The binary format provides generally the same information as the decimal ASCII format but in a compact ...

Страница 8: ...t on EEPROM by keying in the quote character Command mode is required for access to this command 9 A command will change the output format to binary This change is made non volatile in the unit on EEPROM by keying in the quote character Command mode is required for access to this command 10 A command will toggle the output to send a frame of data upon receiving any non command character On to Off ...

Страница 9: ...ational amplifier Each analog output has a 300 ohm resistor in series to eliminate oscillations from high capacitance loads The output range for all of the analog output channels is 5 Volts with respect to the common signal ground The analog outputs can be found on the 12 pin connector The outputs include Signal Pin Range Output Range 0 VDC Scale Factor Bank X Angle 9 30 5V 0 6 0 V Elevation Y ang...

Страница 10: ... Storage 55 C to 85 C Vibration Operating 5g rms 20 Hz to 2 KHz Vibration Survival 10g rms 20 Hz to 2 KHz Shock Survival 500g 1mS sine wave Electrical Frame Rate 250 Hz Maximum Startup Time Data 5 sec Startup Time Fully operational 10 sec Input Power 8 to 45VDC 0 5W Input Current 40mA 12VDC 20mA 24VDC Digital Output RS 232 Analog Output 5VDC Analog Output Impedance 300 Ohm Per line Physical Axis A...

Страница 11: ...he signal ground Dimensions X Y X Z Z Y X Y WATSON INDUSTRIES INC www watson gyro com MADE IN U S A P N Rev Pwr S F Figure 2 12 Pin Connector Pin Description 1 RS 232 TXD 2 Y Angular Rate Analog Output 3 Elevation Y Angle Analog Output 4 No Connection 5 Built in Test BIT Output 6 Power Ground 7 RS 232 RXD 8 Signal Ground 9 X Angular Rate Analog Output 10 Bank X Angle Analog Output 11 Factory use D...

Страница 12: ... including but not limited to any implied warranty of merchantability or fitness for a particular purpose Excluded from any warranty given by Watson Industries are products that have been subject to abuse misuse damage or accident that have been connected installed or adjusted contrary to the instructions furnished by seller or that have been repaired by persons not authorized by Watson Industries...

Страница 13: ... Melby Street Eau Claire WI 54703 ATTN Service Department Telephone 715 839 0628 Fax 715 839 8248 email support watson gyro com Returning the Product Product shall be packaged making sure there is adequate packing around all sides Correspondence shall include Customer s Name and Address Contact Information Equipment Model Number Equipment Serial Number Description of Fault It is the customer s res...

Страница 14: ... 3 ASCII chars representing Octal digits 1 byte Status Bits ST 3 ASCII chars representing Octal digits 1 byte Built In Test BIT Status Bits BT 3 ASCII chars representing Octal digits 1 byte The Flag Bits contain the following information Bit Description If Set 0 Bank Error Flag 1 Elevation Error Flag 2 Unit Temp Out of Range Flag 3 BIT Test Flag 4 X Rate Out of Range Flag 5 Y Rate Out of Range Fla...

Страница 15: ...de active Bit 2 Bit 1 Bit 0 2 The Current System Time Constant 2 Flag Value 1 2 2 1 2 3 8 seconds The Built In Test BIT Status Bits contain the following information Bit Description 0 If set X Accelerometer Sensor Failure 1 If set Y Accelerometer Sensor Failure 2 If set Z Accelerometer Sensor Failure 3 If set Sensor Temperature out of range 4 If set X Angular Rate Sensor Failure 5 If set X Angular...

Страница 16: ...ace bar twice within close succession Note The sensor initialization time begins after the startup message is transmitted Typically this time period lasts 5 seconds but could be as long as 127 seconds depending on your sensor 6 Wait until after initialization is finished At this point command mode will be activated and the unit will accept keyboard commands Note Most Watson sensors have initializa...

Страница 17: ...itting or updating the new channels to restart the data flow Appendix B 3 List Output Channels 1 To list the current output channel selection type 3 at the main menu 2 The sensor will show which channels are currently active Appendix B 4 Select Analog Attitude Full Scale Range 1 To update the scale factor for the analog Bank output type 4 at the main menu This will bring up the menu 2 To change th...

Страница 18: ...1 To change the Elevation Angle to Scale Factor 9 0 V type 2 To change the Elevation Angle to Scale Factor 12 V type 3 To change the Elevation Angle to Scale Factor 18 V type 4 To change the Elevation Angle to Scale Factor 36 V type 5 THE CURRENT SETTING IN DEG V IS 6 0 TYPE IN THE NUMBER OF YOUR SELECTION OR Q TO QUIT 1 BANK ANGLE SCALE FACTOR 6 0 DEG V 2 BANK ANGLE SCALE FACTOR 9 0 DEG V 3 BANK ...

Страница 19: ... immediately To make the new baud rate the default setting send a quote symbol using the new baud rate Otherwise the old baud rate will return the next time the unit is powered up Note The baud rate of the software that is communicating to the sensor will need to be changed to match the new setting 8 Press the space bar to resume data transmission The factory settings for serial communications are...

Страница 20: ...plus 6 bits of data Bit 7 of the LSB is set to 1 Bits 6 to 0 of the LSB are the lower 7 bits of the data As the data words are received the LSB is shifted left to shift out the 7th bit always set to 1 The MSB is then connected to the LSB as a 16 bit word This word is then shifted left to shift out the 7th bit always set to 1 What remains is a signed fractional word with a resolution of 13 bits plu...

Страница 21: ...en subtract 65536 from the data Bank Elevation X Rate Y Rate 2728 912 1668 508 Multiply the data in counts by the scale factor Convert to units counts scale factor For Bank Elevation Scale factor is 180 32768 counts For Angular Rate Scale factor is 200 s 32768 counts Bank Elevation X Rate Y Rate 14 991 5 010 10 181 s 3 100 s The nominal interface settings are 9600 Baud 8 Bit Data 1 Start Bit 1 Sto...

Отзывы: