4-42
Maintenance Procedures
Cleaning the ion block assembly
Clean the ion block assembly if cleaning the sample cone, cone gas nozzle, and
extraction cone fails to increase signal sensitivity.
Disassembling the source ion block assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
O-ring removal kit
•
Needle-nose pliers
To disassemble the ion block assembly
1. Remove the ion block assembly from the source assembly (see
“Removing the ion block assembly from the source assembly” on
).
2. Ensure that the isolation valve is closed.
Warning:
The ion block assembly can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves to perform this procedure.
Isolation valve handle
in closed position
Sampling cone
assembly retaining
blocks
Cone gas nozzle
handle