Cleaning the extraction cone
4-35
Cleaning the extraction cone
Clean the ion block and extraction cone when cleaning the sample cone and
cone gas nozzle fails to increase signal sensitivity. You must remove the ion
block assembly from the source assembly to clean the extraction cone.
Removing the ion block assembly from the source assembly
Required materials
•
Chemical-resistant, powder-free gloves
•
Combined 2.5-mm Allen wrench and cone extraction tool
To remove the ion block assembly
1. Vent and shut down the mass spectrometer (see the mass spectrometer’s
online Help for details)
.
2. Close the source isolation valve (see
“To close the source isolation valve
before starting a maintenance procedure” on page 4-11
Warning:
The source components can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
The source can be hot. To avoid burn injuries, allow it
to cool for at least 30 minutes before proceeding.
Warning:
To avoid puncture wounds, take great care while
working with the source enclosure open, because the ESI probe
tip is sharp.