Replacing the ion block source heater
4-71
Replacing the ion block source heater
Replace the ion block source heater if it fails to heat when the instrument is
pumped down (evacuated).
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
Combined 2.5-mm Allen wrench and cone extraction tool
•
Ion block source heater
To replace the ion block source heater
1. Remove the ion block assembly from the instrument (see
ion block assembly from the source assembly” on page 4-35
).
2. Ensure that the isolation valve is closed.
Warning:
The ion block assembly can be contaminated with
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Isolation valve handle
in closed position